Abstract
OCT (optical coherence tomography) is widely used in ophthalmology and pediatric ophthalmology, but limited research has been done on the use of OCT in strabismus. This study investigates the use of different OCT machines to image rectus muscle insertions pre-, intra-, and post-operatively in pediatric strabismus patients. The OCT machines used in the study were a Bioptigen (Leica Microsystems Inc., Buffalo Grove, IL, USA), Spectralis HRA+OCT with Anterior Segment Module (Heidelberg Engineering, Heidelberg, Germany), Visante (Carl Zeiss, Oberkochen, Germany), and Zeiss Rescan 700 (Carl Zeiss, Oberkochen, Germany). Measurements from the machines were compared with the caliper distance measured during the strabismus surgery before disinsertion or after reattachment. The OCT machines had moderate (Bioptigen: 0.62) to good intraclass correlation coefficients (Rescan: 0.83, Spectralis: 0.85, Visante: 0.88) with intra-operative measurements. To our knowledge, this is the first study to use an operating microscope with integrated intra-operative OCT to image rectus muscle insertions. OCT is a useful tool in strabismus surgical patients in the pre-, intra-, and post-operative settings, particularly in patients who have had previous surgery, when the muscle insertion is unknown. The ability to accurately image rectus muscle insertions has significant implications for the management of strabismus patients.