Affiliation:
1. Hubei Key Laboratory of Modern Manufacturing Quantity Engineering, School of Mechanical Engineering, Hubei University of Technology, Wuhan 430068, China
Abstract
The application of traditional pressure sensors in health monitoring is limited by their initial rigidity. Flexible pressure sensors have thus received extensive attention owing to their excellent device flexibility. In this paper, we demonstrate a method of constructing flexible pressure sensors by inverting porous skeleton nickel foam based on multi-wall carbon nanotubes (MWCNTs) and polydimethylsiloxane (PDMS). MWCNTs and PDMS were mixed to form a composite conductive film, and the mass fraction of MWCNTs was optimized by evaluating the resistance change rate of the composite film. The optimized value of the mass fraction was 5%, which was used to prepare the flexible pressure sensors. The response, hysteresis, and stability of the sensors were further characterized. Pulse signals of humans were detected through flexible sensors, which can be used to evaluate cardiovascular conditions of the human body. These performance characteristics and the application demonstration show that our flexible pressure sensors have good prospects in human health care.
Funder
National Natural Science Foundation of China
Hubei Provincial Natural Science Foundation of China
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献