Abstract
Electro-wetting-on-dielectric (EWOD) enables the manipulation of droplets on a two-dimensional surface, which provides a versatile technique for digital microfluidics at a micro- or nano-scale. However, the deficiency of the dispensing precision has long limited its applications in micro total analysis systems (μ-TAS) where the accuracy of assays is largely determined by the volume control of the reagent dosing. This paper proposes optimum electrode designs and carries out characterization experiments to demonstrate the reproducibility of on-chip droplet generation with no extra external apparatus. The coefficient variation of the volumes of consecutively dispensed droplets from a non-refilling reservoir can be limited to below 0.3%, indicating the validity of the new electrode structure in practical applications.
Funder
Fundamental Research Funds for the Central Universities
State Key Laboratory of ASIC and System, Fudan University
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Control and Systems Engineering
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献