We report an absolute interferometer configured with a 1 GHz microwave source photonically synthesized from a fiber mode-locked laser of a 100 MHz pulse repetition rate. Special attention is paid to the identification of the repeatable systematic error with its subsequent suppression by means of passive compensation as well as active correction. Experimental results show that passive compensation permits the measurement error to be less than 7.8 μm (1 σ) over a 2 m range, which further reduces to 3.5 μm (1 σ) by active correction as it is limited ultimately by the phase-resolving power of the phasemeter employed in this study. With precise absolute distance ranging capability, the proposed scheme of the photonic microwave interferometer is expected to replace conventional incremental-type interferometers in diverse long-distance measurement applications, particularly for large machine axis control, precision geodetic surveying and inter-satellite ranging in space.