Abstract
The time-base used by digital storage oscilloscopes allows limited selections of the sample rate, namely constrained to a few integer submultiples of the maximum sample rate. This limitation offers the advantage of simplifying the data transfer from the analog-to-digital converter to the acquisition memory, and of assuring stability performances, expressed in terms of absolute jitter, that are independent of the chosen sample rate. On the counterpart, it prevents an optimal usage of the memory resources of the oscilloscope and compels to post processing operations in several applications. A time-base that allows selecting the sample rate with very fine frequency resolution, in particular as a rational submultiple of the maximum rate, is proposed. The proposal addresses the oscilloscopes with time-interleaved converters, that require a dedicated and multifaceted approach with respect to architectures where a single monolithic converter is in charge of signal digitization. The proposed time-base allows selecting with fine frequency resolution sample rate values up to 200 GHz and beyond, still assuring jitter performances independent of the sample rate selection.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献