Author:
Kim Pilsu,Kang Jeeun,Song and Tai-Kyong
Abstract
The implementation of dynamic delay calculations (DDCs) is challenging for ultra-compact ultrasound imaging due to the enormous computation and power consumption requirements. Here, we present an efficient pseudo-DDC method based on optimal zone segmentation (PDC-Optimal), which significantly decreases these requirements relative to an unconstrained DDC method: reductions in flip-flops of 84.35% and in look-up tables of 94.19%, respectively. The reductions lead to an up to 94.53% lower dynamic power consumption and provide image quality comparable to the unconstrained DDC method. The proposed PDC-Optimal method also provides adaptive flexibility between beamforming accuracy and battery life using the delay error allowance, a user-definable parameter. A conventional pseudo-DDC method using uniform zone segmentation (PDC-Conv) presented substantial image degradation in the near imaging field when the same number of zone segments was used. Therefore, the PDC-Optimal method provides an efficient yet flexible DDC solution to improve the experiences for ultra-compact ultrasound imaging system users.
Funder
National Research Foundation of Korea
Subject
Electrical and Electronic Engineering,Computer Networks and Communications,Hardware and Architecture,Signal Processing,Control and Systems Engineering
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献