Abstract
Ion selective electrodes (ISE) were evaluated for use in a flue gas desulfurization (FGD) wastewater monitoring system. Calcium, chloride, nitrate, and nitrite ISE’s were calibrated in a broad range of concentrations that were designed to model an actual FGD wastewater sample that was obtained from an industrial partner. Ideal Nernst, Modified Nernst, and a multiparameter regression analysis was performed for each electrode and evaluated on general fit and sensitivity at anticipated operating conditions. The Ideal Nernst equation, even with correction for ionic strength, was not able to properly model actual electrode performance. The multiparameter regression was able to model the electrode performance with relative errors of 10–25% when ionic strengths were below 0.1 M. Through the evaluation of real-time sensor usage at real conditions, a methodology of washing and sampling rate is suggested to minimize error in the readings.
Funder
Ohio Coal Development Office
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献