The paper presents the results of in vitro studies of fretting and fretting corrosion processes of Ti6Al4V implant alloy in the environment of natural saliva and self-made mucin-based artificial saliva solutions. The study was performed on a specially designed fretting pin-on-disc tester, which was combined with a set used for electrochemical research. The open circuit potential measurements and potentiodynamic method were used for corrosion tests. The worn surfaces were subjected to microscopic observations and an evaluation of wear. Results were interpreted using the dissipated energy and third-body approaches. The X-ray diffraction analysis showed that titanium oxides constitute over 80% of the friction products. Special attention was paid to the role of saliva and its substitutes, which in certain cases can lead to the intensification of fretting wear. On the basis of the received results, a new phenomenological model of fretting corrosion processes was proposed. This model involves the formation of an abrasive paste that is a combination of metal oxides and the organic components of saliva.