Author:
Zhang Hualin,Grams Michael P.,Foy Joseph J.,Mayr Nina A.
Abstract
Computations of heterogeneity dose parameters in GRID therapy remain challenging in many treatment planning systems (TPS). To address this difficulty, we developed reference dose tables for a standard GRID collimator and validate their accuracy. The .decimal Inc. GRID collimator was implemented within the Eclipse TPS. The accuracy of the dose calculation was confirmed in the commissioning process. Representative sets of simulated ellipsoidal tumours ranging from 6–20 cm in diameter at a 3-cm depth; 16-cm ellipsoidal tumours at 3, 6, and 10 cm in depth were studied. All were treated with 6MV photons to a 20 Gy prescription dose at the tumour center. From these, the GRID therapy dosimetric parameters (previously recommended by the Radiosurgery Society white paper) were derived. Differences in D5 through D95 and EUD between different tumour sizes at the same depth were within 5% of the prescription dose. PVDR from profile measurements at the tumour center differed from D10/D90, but D10/D90 variations for the same tumour depths were within 11%. Three approximation equations were developed for calculating EUDs of different prescription doses for three radiosensitivity levels for 3-cm deep tumours. Dosimetric parameters were consistent and predictable across tumour sizes and depths. Our study results support the use of the developed tables as a reference tool for GRID therapy.
Cited by
12 articles.
订阅此论文施引文献