Fabrication and Analysis of a Ti6Al4V Implant for Cranial Restoration

Author:

Moiduddin KhajaORCID,Mian Syed HammadORCID,Umer UsamaORCID,Alkhalefah Hisham

Abstract

A custom made implant is critical in cranioplasty to cushion and restore intracranial anatomy, as well as to recover the appearance and attain cognitive stability in the patient. The utilization of customized titanium alloy implants using three-dimensional (3D) reconstruction technique and fabricated using Electron Beam Melting (EBM) has gained significant recognition in recent years, owing to their convenience and effectiveness. Besides, the conventional technique or the extant practice of transforming the standard plates is unreliable, arduous and tedious. As a result, this work aims to produce a customized cranial implant using 3D reconstruction that is reliable in terms of fitting accuracy, appearance, mechanical strength, and consistent material composition. A well-defined methodology initiating from EBM fabrication to final validation has been outlined in this work. The custom design of the implant was carried out by mirror reconstruction of the skull’s defective region, acquired through computer tomography. The design of the customized implant was then analyzed for mechanical stresses by applying finite element analysis. Consequently, the 3D model of the implant was fabricated from Ti6Al4V ELI powder with a thickness of ≃1.76–2 mm. Different tests were employed to evaluate the bio-mechanical stability and strength of the fabricated customized implant design. A 3D comparison study was performed to ensure there was anatomical accuracy, as well as to maintain gratifying aesthetics. The bio-mechanical analysis results revealed that the maximum Von Mises stress (2.5 MPa), strain distribution (1.49 × 10−4) and deformation (3.26 × 10−6 mm) were significantly low in magnitude, thus proving the implant load resistance ability. The average yield and tensile strengths for the fabricated Ti6Al4V ELI EBM specimen were found to be 825 MPa and 880 MPa, respectively, which were well over the prescribed strength for Ti6Al4V ELI implant material. The hardness study also resulted in an acceptable outcome within the acceptable range of 30–35 HRC. Certainly, the chemical composition of the fabricated EBM specimen was intact as established in EDX analysis. The weight of the cranial implant (128 grams) was also in agreement with substituted defected bone portion, ruling out any stress shielding effect. With the proposed approach, the anatomy of the cranium deformities can be retrieved effectively and efficiently. The implementation of 3D reconstruction techniques can conveniently reduce tedious alterations in the implant design and subsequent errors. It can be a valuable and reliable approach to enhance implant fitting, stability, and strength.

Funder

Deanship of scientific research, King Saud University

Publisher

MDPI AG

Subject

Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science

Cited by 26 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3