Abstract
This paper presents a specifically designed grating-corner-cube sensor for precise roll angle measurements. Owing to the diffraction characteristics of the transmission grating and reflection characteristics of the corner cube, two spatially separated parallel beams are naturally constructed. Through differential detection of the positions of two parallel beams, we experimentally demonstrate the possibility of a precise roll angle measurement at a high refresh rate. A performance evaluation of the proposed technique indicates a stability of 0.46 arcsec over 5 min. Compared with a commercial autocollimator over a range of 500 arcsec, the residuals are maintained within ±2 arcsec with a standard deviation of 1.37 arcsec. Furthermore, a resolution of 0.8 arcsec can be achieved using the proposed method. The developed compact roll angle sensor has potential applications in academic and industrial fields.
Funder
Natural Science Foundation of Beijing Municipality
National Natural Science Foundation of China
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献