Affiliation:
1. Department of Cardiac Surgery, Institute for Clinical and Experimental Medicine, Prague, Czech Republic. onsz@ikem.cz
Abstract
Mechanical circulatory support (MCS) with an implantable left ventricular assist device (LVAD) is an established therapeutic option for advanced heart failure. Most of the currently used LVADs generate a continuous stream of blood that decreases arterial pulse pressure. This study investigated whether a change of the pulse pressure during different pump speed settings would affect cerebral autoregulation and thereby affect cerebral blood flow (CBF). The study included 21 haemodynamically stable outpatients with a continuous-flow LVAD (HeartMate II, Abbott, USA) implanted a median of 6 months before the study (interquartile range 3 to 14 months). Arterial blood pressure (measured by finger plethysmography) was recorded simultaneously with CBF (measured by transcranial Doppler ultrasound) during baseline pump speed (8900 rpm [IQR 8800; 9200]) and during minimum and maximum tolerated pump speeds (8000 rpm [IQR 8000; 8200] and 9800 rpm [IQR 9800; 10 000]). An increase in LVAD pump speed by 800 rpm [IQR 800; 1000] from the baseline lead to a significant decrease in arterial pulse pressure and cerebral blood flow pulsatility (relative change −24 % and −32 %, both p < 0.01), but it did not affect mean arterial pressure and mean CBF velocity (relative change 1 % and −1.7 %, p=0.1 and 0.7). In stable patients with a continuous-flow LVAD, changes of pump speed settings within a clinically used range did not impair static cerebral autoregulation and cerebral blood flow.
Publisher
Institute of Physiology of the Czech Academy of Sciences
Subject
General Medicine,Physiology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献