Reverse Endoventricular Artificial Obturator in Tricuspid Valve Position. Experimental Feasibility Research Study
-
Published:2014-04-16
Issue:
Volume:
Page:157-165
-
ISSN:1802-9973
-
Container-title:Physiological Research
-
language:en
-
Short-container-title:Physiol Res
Author:
SOCHMAN J.1, PEREGRIN J. H., PAVCNIK D., UCHIDA B. T., TIMMERMANS H. A., SEDMERA D., BENADA O., KOFRONOVA O., KELLER F. S., ROSCH J.
Affiliation:
1. Clinic of Cardiology, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
Abstract
The concept of vena contracta space reduction in tricuspid valve position was tested in an animal model. Feasibility of specific artificial obturator body (REMOT) fixed to the right ventricular apex and interacting with tricuspid valve leaflets was evaluated in three different animal studies. Catheter-based technique was used in three series of experiment in 7 sheep. First acute study was designed for evaluation if the screwing mode of guide wire anchoring to the right ventricular apex is feasible for the whole REMOT body fixing. Longer study was aimed to evaluate stability of the REMOT body in desired position when fixing the screwing wire on its both ends (to the right ventricular apex and to the skin in the neck area). X-ray methods and various morphological methods were used. The third acute study was intended to the REMOT body deployment without any fixing wire. In all of 7 sheep the REMOT was successfully inserted into the right heart cavities and then fixed to the right ventricular apex area. When the REMOT was left in situ more than 6 months it was stable, induced adhesion to the tricuspid valve leaflet and was associated with a specific cell invasion. Releasing of the REMOT from the guiding tools was also successfully verified. Deployment of the obturator body in the aim to reduce the tricuspid valve orifice is feasible and well tolerated in the short and longer term animal model. Specific cell colonization including neovascularization of the obturator body was observed.
Publisher
Institute of Physiology of the Czech Academy of Sciences
Subject
General Medicine,Physiology
Reference24 articles.
1. BENES J, KAZDOVA L, DRAHOTA Z, HOUSTEK J, MEDRIKOVA D, KOPECKY J, KOVAROVA N, VRBACKY M, SEDMERA D, STRNAD H, KOLAR M, PETRAK J, BENADA O, SKAROUPKOVA P, CERVENKA, L, MELENOVSKY V: Effect of metformin therapy on cardiac function and survival in a volume-overload model of heart failure in rats. Clin Sci 121: 29-41, 2011. 2. CHAN KMJ, ZAKKAR M, AMIRAK E, PUNJABI PP: Tricuspid valve disease: pathophysiology and optimal management. Progress Cardiovasc Dis 51: 482-486, 2009. 3. CHIAM PTL, RUIZ CE: Percutaneous transcatheter mitral valve repair. A classification of the technology. JACC Cardiovasc Interv 4: 1-13, 2011. 4. DE BONIS M, LAPENA E, LA CANNA G, GRIMALDI A, MAISANO F, TORRACA F, CALDAROLA A, ALFIERI O: A novel technique for correction of severe tricuspid valve regurgitation due to complex lesions. Eur J Cardiothoracic Surg 25: 760-765, 2004. 5. FORFIA PR, FISHER MR, MATYSI SC, HOUSTEN-HARRIS T, HEMNES AR, BORLAUG BA, CHAMERA E, CORRETTI MC, CHAMPION HS, ABRAHAM TP, BORGIS RE, HASSOUN PM: Tricuspid annular displacement predicts survival in pulmonary hypertension. Am J Respir Crit Care Med 174: 1034-1041, 2006.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|