Affiliation:
1. Universidade Federal de Minas Gerais, Faculdade de Farmácia, Departamento de Produtos Farmacêuticos, Av. Pres. Antônio Carlos 6627, 31270-901, Belo Horizonte, MG, Brazil
Abstract
Abstract
Tuberculosis treatment consists of a fixed dose combination of rifampicin (RIF), isoniazid (INH), pyrazinamide (PYZ), and ethambutol hydrochloride (EMB). The combined treatment using various drugs is necessary for patient curing, without recrudescence, and for prevention of drug-resistant mutants, which may occur during treatment. An HPLC-diode array detector (DAD) method for the simultaneous determination of RIF, INH, PYZ, and EMB in fixed dose combination tablets was developed and validated. Chromatographic experiments were performed on an Agilent 1200 HPLC system, and the separation was carried out on a Purospher STAR RP18e (250 × 4.6 mm id, 5 μm, Merck) analytical column. Gradient elution was carried out with a mobile phase of 20 mM monobasic sodium phosphate buffer with 0.2% triethylamine (pH 7.0) and acetonitrile at a flow rate of 1.5 mL/min. The total run time was 12 min, and the re-equilibration time was 5 min. EMB detection was performed at 210 nm, and RIF, INH, and PYZ were detected at 238 nm, using a DAD. The method proved to be specific, linear (r2 > 0.99), precise (RSD <2%), accurate, and robust and may be applied to the QC analysis of pharmaceutical formulations.
Publisher
Oxford University Press (OUP)
Subject
Pharmacology,Agronomy and Crop Science,Environmental Chemistry,Food Science,Analytical Chemistry
Cited by
30 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献