Abstract
Abstract
Using United States Pharmacopeia-National Formulary (USP-NF) general method <1223> guidance, the Soleris® automated system and reagents (Nonfermenting Total Viable Count for bacteria and Direct Yeast and Mold for yeast and mold) were validated, using a performance equivalence approach, as an alternative to plate counting for total microbial content analysis using five representative microbes: Staphylococcus aureus, Bacillus subtilis, Pseudomonas aeruginosa, Candida albicans, and Aspergillus brasiliensis. Detection times (DTs) in the alternative automated system were linearly correlated to CFU/sample (R2 = 0.94–0.97) with ≥70% accuracy per USP General Chapter <1223> guidance. The LOD and LOQ of the automated system were statistically similar to the traditional plate count method. This system was significantly more precise than plate counting (RSD 1.2–2.9% for DT, 7.8–40.6% for plate counts), was statistically comparable to plate counting with respect to variations in analyst, vial lots, and instruments, and was robust when variations in the operating detection thresholds (dTs; ±2 units) were used. The automated system produced accurate results, was more precise and less labor-intensive, and met or exceeded criteria for a valid alternative quantitative method, consistent with USP-NF general method <1223> guidance.