Effect of metal implants and metal artifacts on back‐projected two‐dimensional entrance fluence determined by EPID dosimetry

Author:

Cao Zheng12ORCID,Gao Xiang2ORCID,Liu Gongfa1,Pei Yuanji1ORCID

Affiliation:

1. National Synchrotron Radiation Laboratory University of Science and Technology of China Hefei China

2. Hematology & Oncology Department Hefei First People's Hospital Hefei China

Abstract

AbstractPurposeTo evaluate the errors caused by metal implants and metal artifacts in the two‐dimensional entrance fluences reconstructed using the back‐projection algorithm based on electronic portal imaging device (EPID) images.MethodsThe EPID in the Varian VitalBeam accelerator was used to acquire portal dose images (PDIs), and then commercial EPID dosimetry software was employed to reconstruct the two‐dimensional entrance fluences based on computed tomography (CT) images of the head phantoms containing interchangeable metal‐free/titanium/aluminum round bars. The metal‐induced errors in the two‐dimensional entrance fluences were evaluated by comparing the γ results and the pixel value errors in the metal‐affected regions. We obtained metal‐artifact‐free CT images by replacing the voxel values of non‐metal inserts with those of metal inserts in metal‐free CT images to evaluate the metal‐artifact‐induced errors.ResultsThe γ passing rates (versus PDIs obtained without a phantom in the beam field (PDIair), 2%/2 mm) for the back‐projected two‐dimensional entrance fluences of phantoms containing titanium or aluminum (BPTi/BPAl) were reduced from 92.4% to 90.5% and 90.6%, respectively, relative to the metal‐free phantom (BPmetal‐free). Titanium causes more severe metal artifacts in CT images than aluminum, and its removal resulted in a 0.0022 CU (median) reduction in the pixel value of BPTi artifact‐free relative to BPTi in the metal‐affected region. Moreover, the mean absolute error (MAE) and root mean square error (RMSE) decreased from 0.0050 CU and 0.0063 CU to 0.0034 CU and 0.0040 CU, respectively (vs. BPmetal‐free).ConclusionMetal implants increase the errors in back‐projected two‐dimensional entrance fluences, and metals with higher electron densities cause more errors. For high‐electron‐density metal implants that produce severe metal artifacts (e.g., titanium), removing metal artifacts from the CT images can improve the accuracy of the two‐dimensional entrance fluences reconstructed by back‐projection algorithms.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3