Evaluation and improvement of angular response for a commercial 2D detector array for patient‐specific QA

Author:

Wang Sha1,Zheng Yuanshui1

Affiliation:

1. Guangzhou Concord Cancer Center Sino‐Singapore Guangzhou Knowledge City Guangzhou City China

Abstract

AbstractPurposeMatriXX ionization chamber array has been widely used for the composite dose verification of IMRT/VMAT plans. However, in addition to its dose response dependence on gantry angle, there seems to be an offset between the beam axis and measured dose profile by MatriXX for oblique beam incidence at various gantry angles, leading to unnecessary quality assurance (QA) fails. In this study, we investigated the offset at various setup conditions and how to eliminate or decrease it to improve the accuracy of MatriXX for IMRT/VMAT plan verification with original gantry angles.MethodsWe measured profiles for a narrow beam with MatriXX located at various depths in increments of 0.5 mm from the top to bottom of the sensitive volume of the array detectors and gantry angles from 0° to 360°. The optimal depth for QA measurement was determined at the depth where the measured profile had minimum offset.ResultsThe measured beam profile offset varies with incident gantry angle, increasing from vertical direction to lateral direction, and could be over 3 cm at vendor‐recommended depth for near lateral direction beams. The offset also varies with depth, and the minimum offset (almost 0 for most oblique beams) was found to be at a depth of ∼2.5 mm below the vendor suggested depth, which was chosen as the optimal depth for all QA measurements. Using the optimal depth we determined, QA results (3%/2 mm Gamma analysis) were largely improved with an average of 99.4% gamma passing rate (no fails for 95% criteria) for 10 IMRT and VMAT plans with original gantry angles compared to 94.1% using the vendor recommended depth.ConclusionsThe improved accuracy and passing rate for QA measurement performed at the optimal depth with original gantry angles would lead to reduction in unnecessary repeated QA or plan changes due to QA system errors.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3