Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT

Author:

Hernández Katerine Viviana Díaz1,Unterkirhers Sergejs2,Schneider Uwe12

Affiliation:

1. Science Faculty University of Zürich Zürich Switzerland

2. Medical Physics Radiotherapy Hirslanden Zürich Switzerland

Abstract

AbstractPurposeThe purpose of this study is to assess the quality of automatic planned O‐Ring Halcyon linac SBRT plans for pelvic lymph node metastases and to establish an absolute PTV volume threshold as a plan quality prediction criterion. Compliance of the plans to institutional SBRT plan evaluation criteria and differences in plan quality and treatment delivery times between Halcyon Linac and CyberKnife robotic SBRT were evaluated.MethodsTwenty‐one CyberKnife treatment plans were replanned for Halcyon. Prescription doses range was 26–40 Gy in mean three fractions. The mean/median planning target volume was 4.0/3.6 cm3. Institutional criteria for the plan evaluation were: New Conformity Index (NCI), Conformity Index (CI), Modified Gradient Index (MGI), selectivity index reciprocal (PIV/TVPIV), and the target coverage by prescription isodose (%PIV). Statistical analysis based on the receiver operating characteristic (ROC) curve was used to determine a plan quality predictor threshold of the PTV volume. Comparative analysis of normal tissue complication probabilities (NTCP) was used to assess the risk of toxicity in healthy tissues.ResultsSeventy‐one percent (n = 15)/95% (n = 20) of Halcyon and 81% (n = 17)/100% (n = 21) of CK plans fulfilled all ideal/tolerance criteria. For PTVs above a found optimal threshold of 2.6 cm3 (71%, n = 15), no statistically significant difference was observed between the CI, NCI, PIV/TVPIV, and MGI indexes of both groups, while the coverage (%PIV) was statistically but not clinically significantly different between cohorts. Significantly shorter delivery times are expected with Halcyon. No significant differences in NTCP were observed.ConclusionAll but one automatically optimized Halcyon treatment plans demonstrated ideal or acceptable performance. PTV threshold of 2.6 cm3 can be used as decision criteria in clinical settings. The results of our study demonstrated the promising performance of the Halcyon for pelvic SBRT, although plan‐specific QA is required to verify machine performance during plan delivery.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3