Monte Carlo investigation of dose distribution of uniformly and non‐uniformly loaded standard and notched eye plaques

Author:

Semeniuk Oleksii1,Malkov Victor2ORCID,Chamberland Marc J. P.3,Weersink Robert A.24

Affiliation:

1. Radiation Medicine Program Princess Margaret Cancer Center Toronto Canada

2. Department of Radiation Oncology University of Toronto Toronto Canada

3. Department of Radiation Oncology University of Vermont Medical Center Vermont USA

4. Department of Medical Biophysics University of Toronto Toronto Canada

Abstract

AbstractTo investigate the effect of using non‐uniform loading and notched plaques on dose distribution for eye plaques.Using EGSnrc Monte Carlo (MC) simulations, we investigate eye plaque dose distributions in water and in an anatomically representative eye phantom. Simulations were performed in accordance with TG43 formalism and compared against full MC simulations which account for inter‐seed and inhomogeneity effects.For standard plaque configurations, uniformly and non‐uniformly loaded plaque dose distributions in water showed virtually no difference between each other. For standard plaque, the MC calculated dose distribution in planes parallel to the plaque is narrower than the TG43 calculation due to attenuation at the periphery of the plaque by the modulay. MC calculated the dose behind the plaque is fully attenuated. Similar results were found for the notched plaque, with asymmetric attenuation along the plane of the notch. Cumulative dose volume histograms showed significant reductions in the calculated MC doses for both tumor and eye structures, compared to TG43 calculations. The effect was most pronounced for the notch plaque where the MC dose to the optic nerve was greatly attenuated by the modulay surrounding the optic nerve compared to the TG43. Thus, a reduction of optic nerve D95% from 14 to 0.2 Gy was observed, when comparing the TG43 calculation to the MC result. The tumor D95% reduced from 89.2 to 79.95 Gy for TG43 and MC calculations, respectively.TG43 calculations overestimate the absolute dose and the lateral dose distribution of both standard and notched eye plaques, leading to the dose overestimation for the target and organs at risk. The dose matching along the central axis for the non‐uniformly loaded plaques to that of uniformly loaded ones was found to be sufficient for providing comparable coverage and can be clinically used in eye‐cancer‐busy centers.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3