A scintillation dosimeter with real‐time positional tracking information for in vivo dosimetry error detection in HDR brachytherapy

Author:

Tho Daline1ORCID,Lavallée Marie‐Claude23,Beaulieu Luc23

Affiliation:

1. Department of Radiation Oncology The University of Texas MD Anderson Cancer Center Houston Texas USA

2. Département de physique, de génie physique et d'optique, et Centre de recherche sur le cancer Université Laval Québec Québec Canada

3. Service de physique médicale et de radioprotection, Centre intégré de cancérologie CHU de Québec‐Université Laval et Centre de recherche du CHU de Québec Québec Canada

Abstract

AbstractPurposeTo evaluate the performance of an electromagnetic (EM)‐tracked scintillation dosimeter in detecting source positional errors of IVD in HDR brachytherapy treatment.Materials and MethodsTwo different scintillator dosimeter prototypes were coupled to 5 degrees‐of‐freedom (DOF) EM sensors read by an Aurora V3 system. The scintillators used were a 0.3 × 0.4 × 0.4 mm3 ZnSe:O and a BCF‐60 plastic scintillator of 0.5 mm diameter and 2.0 mm in length (Saint‐Gobain Crystals). The sensors were placed at the dosimeter's tip at 20.0 mm from the scintillator. The EM sampling rate was 40/s while the scintillator signal was sampled at 100 000/s using two photomultiplier tubes from Hamamatsu (series H10722) connected to a data acquisition board. A high‐pass filter and a low‐pass filter were used to separate the light signal into two different channels. All measurements were performed with an afterloader unit (Flexitron‐Elekta AB, Sweden) in full‐scattered (TG43) conditions. EM tracking was further used to provide distance/angle‐dependent energy correction for the ZnSe:O inorganic scintillator. For the error detection part, lateral shifts of 0.5 to 3 mm were induced by moving the source away from its planned position. Indexer length (longitudinal) errors between 0.5 to 10 mm were also introduced. The measured dose rate difference was converted to a shift distance, with and without using the positional information from the EM sensor.ResultsThe inorganic scintillator had both a signal‐to‐noise‐ratio (SNR) and signal‐to‐background‐ratio (SBR) close to 70 times higher than those of the plastic scintillator. The mean absolute difference from the dose measurement to the dose calculated with TG‐43U1 was 1.5% ±0.7%. The mean absolute error for BCF‐60 detector was 1.7% when compared to TG‐43 calculations formalism. With the inorganic scintillator and EM tracking, a maximum area under the curve (AUC) gain of 24.0% was obtained for a 0.5‐mm lateral shift when using the EMT data with the ZnSe:O. Lower AUC gains were obtained for a 3‐mm lateral shifts with both scintillators. For the plastic scintillator, the highest gain from using EM tracking information occurred for a 0.5‐mm lateral shift at 20 mm from the source. The maximal gain (17.4%) for longitudinal errors was found at the smallest shifts (0.5 mm).ConclusionsThis work demonstrates that integrating EM tracking to in vivo scintillation dosimeters enables the detection of smaller shifts, by decreasing the dosimeter positioning uncertainty. It also serves to perform position‐dependent energy correction for the inorganic scintillator,providing better SNR and SBR, allowing detection of errors at greater distances from the source.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Radioluminescence Dosimetry in Modern Radiation Therapy;Advanced Photonics Research;2024-07-15

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3