Commissioning and implementing a Quality Assurance program for dedicated radiation oncology MRI scanners

Author:

Poulin Eric12ORCID,Lacroix Frederic12,Archambault Louis12,Jutras Jean‐David12

Affiliation:

1. Département de physique de génie physique et d'optique et Centre de recherche sur le cancer de l'Université Laval Université Laval Québec Canada

2. Département de radio‐oncologie et Axe Oncologie du Centre de recherche du CHU de Québec CHU de Québec‐Université Laval Québec Canada

Abstract

AbstractPurposeACR and AAPM task group's guidelines addressing commissioning for dedicated MR simulators were recently published. The goal of the current paper is to present the authors' 2‐year experience regarding the commissioning and introduction of a QA program based on these guidelines and an associated automated workflow.MethodsAll mandatory commissioning tests suggested by AAPM report 284 were performed and results are reported for two MRI scanners (MAGNETOM Sola and Aera). Visual inspection, vendor clinical or service platform, third‐party software, or in‐house python‐based code were used. Automated QA and data analysis was performed via vendor, in‐house or third‐party software. QATrack+ was used for QA data logging and storage. 3D geometric distortion, B0 inhomogeneity, EPI, and parallel imaging performance were evaluated.ResultsContrasting with AAPM report 284 recommendations, homogeneity and RF tests were performed monthly. The QA program allowed us to detect major failures over time (shimming, gradient calibration and RF interference). Automated QA, data analysis, and logging allowed fast ACR analysis daily and monthly QA to be performed in 3 h. On the Sola, the average distortion is 1 mm for imaging radii of 250 mm or less. For radii of up to 200 mm, the maximum, average (standard deviation) distortion is 1.2  and 0.4 mm (0.3 mm). Aera values are roughly double the Sola for radii up to 200 mm. EPI geometric distortion, ghosting ratio, and long‐term stability were found to be under the maximum recommended values. Parallel imaging SNR ratio was stable and close to the theoretical value (ideal g‐factor). No major failures were detected during commissioning.ConclusionAn automated workflow and enhanced QA program allowed to automatically track machine and environmental changes over time and to detect periodic failures and errors that might otherwise have gone unnoticed. The Sola is more geometrically accurate, with a more homogenous B0 field than the Aera.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3