Development of a new gastrointestinal endoscope forceps plug that can minimize gas leakage

Author:

Ishida Tomo12,Hayashi Yoshinori12,Nose Yohei12ORCID,Goto Kohei3,Ishii Yuji4,Saito Takuro2,Yamashita Kotarou2,Tanaka Koji2,Yamamoto Kazuyoshi2,Makino Tomoki2,Takahashi Tsuyoshi2,Kurokawa Yukinori2,Eguchi Hidetoshi2,Doki Yuichiro2,Nakajima Kiyokazu12ORCID

Affiliation:

1. Department of Next Generation Endoscopic Intervention (Project ENGINE) Osaka University Graduate School of Medicine Osaka Japan

2. Department of Gastroenterological Surgery Osaka University Graduate School of Medicine Osaka Japan

3. Department of Endoscope‐related Products, Sales Division TOP Corporation Tokyo Japan

4. Medical and Welfare Development Division Castem Corporation Hiroshima Japan

Abstract

AbstractBackground and aimsIn our previous study, we visualized and systematically evaluated gas leakage from the forceps plug of the gastrointestinal endoscope system using the Schlieren system. In order to minimize the potential risk of infection from gas leakage from the gastrointestinal endoscope, the development of a new forceps plug was considered urgent. In this study, we analyzed the structure of commercially available forceps plugs and attempted to develop new forceps plugs with improvements.MethodsMicrofocus computed tomography was used to nondestructively analyze the structural changes that occur when forceps are inserted into a commercially available forceps plug. Based on the findings, the basic structure of the newly developed forceps plug was set. We examined the airtightness of these newly developed plugs using the Schlieren system and also compared their fractional resistance with those of commercially available plugs.ResultsAs a result of the nondestructive analysis, all of the commercially available plugs had a single valve, and the cleavage created in the valve during forceps, insertion was large in the case of plugs with slit‐type entrances. In the newly developed forceps plugs, all four types of plugs showed less gas leakage and similar or better usability compared to the commercially available plugs.ConclusionsThe structural weaknesses of the existing gastrointestinal endoscopic forceps plugs were identified. Based on the findings, we froze the design of a prototype of a new forceps plug that was airtight and not inferior in usability compared to commercially available plugs.

Publisher

Wiley

Subject

Organic Chemistry,Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3