Irreversible electroporation (IRE) is a new technique for tumour cell ablation that is reported to involve non-thermal-based energy using high voltage at short microsecond pulse lengths. In vivo assessment of the thermal energy generated during IRE has not been performed. Thermal injury can be predicted using a critical temperature model. The aim of this study was to assess the potential for thermal injury during IRE in an in vivo porcine model.
Methods
In vivo continuous temperature assessments of 86 different IRE procedures were performed on porcine liver, pancreas, kidney and retroperitoneal tissue. Tissue temperature was measured continuously throughout IRE by means of two thermocouples placed at set distances (0·5 cm or less, and 1 cm) from the IRE probes within the treatment field. Thermal injury was defined as a tissue temperature of 54°C lasting at least 10 s. Tissue type, pulse length, probe exposure length, number of probes and retreatment were evaluated for associations with thermal injury. In addition, IRE ablation was performed with metal clips or metal stents within the ablation field to determine their effect on thermal injury.
Results
An increase in tissue temperature above the animals' baseline temperature (median 36·0°C) was generated during IRE in all tissues studied, with the greatest increase found at the thermocouple placed within 0·5 cm in all instances. On univariable and multivariable analysis, ablation in kidney tissue (maximum temperature 62·8°C), ablation with a pulse length setting of 100 µs (maximum 54·7°C), probe exposure of at least 3·0 cm (maximum 52·0°C) and ablation with metal within the ablation field (maximum 65·3°C) were all associated with a significant risk of thermal injury.
Conclusion
IRE can generate thermal energy, and even thermal injury, based on tissue type, probe exposure lengths, pulse lengths and proximity to metal. Awareness of probe placement regarding proximity to critical structures as well as probe exposure length and pulse length are necessary to ensure safety and prevent thermal injury. A probe exposure of 2·5 cm or less for liver IRE, and 1·5 cm or less for pancreas, with maximum pulse length of 90 µs will result in safe and non-thermal energy delivery with spacing of 1·5–2·3 cm between probe pairs.
Publisher
Oxford University Press (OUP)
Reference32 articles.
1. Radiofrequency ablation: the experts weigh in;Tanabe;Cancer,2004
2. Radiofrequency ablation for unresectable tumors of the liver;Howard;Am Surg,2008
3. Irreversible electroporation of locally advanced pancreatic head adenocarcinoma;Martin;J Gastrointest Surg,2013
4. Radiofrequency ablation of the pancreas. I: Definition of optimal thermal kinetic parameters and the effect of simulated portal venous circulation in an ex-vivo porcine model;Date;JOP,2005
5. Radiofrequency ablation of porcine liver in vivo: effects of blood flow and treatment time on lesion size;Patterson;Ann Surg,1998
Cited by
97 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献