Reproducibility of absolute quantification of adenosine triphosphate and inorganic phosphate in the liver with localized 31P‐magnetic resonance spectroscopy at 3‐T using different coils

Author:

Jonuscheit Marc12ORCID,Wierichs Stefan12,Rothe Maik123,Korzekwa Benedict12,Mevenkamp Julian4,Bobrov Pavel25,Kupriyanova Yuliya12,Roden Michael126,Schrauwen‐Hinderling Vera B.124

Affiliation:

1. Institute for Clinical Diabetology, German Diabetes Center Leibniz Institute for Diabetes Research at Heinrich Heine University Düsseldorf Düsseldorf Germany

2. German Center for Diabetes Research (DZD e.V.) München‐Neuherberg Germany

3. University Clinic and Outpatient Clinic for Radiology University Hospital Halle (Saale) Halle (Saale) Germany

4. Department of Radiology and Nuclear Medicine Maastricht University Medical Center Maastricht The Netherlands

5. Institute for Biometrics and Epidemiology, German Diabetes Center Leibniz Institute for Diabetes Research at Heinrich Heine University Düsseldorf Düsseldorf Germany

6. Department of Endocrinology and Diabetology, Medical Faculty and University Hospital Düsseldorf Heinrich Heine University Düsseldorf Düsseldorf Germany

Abstract

AbstractConcentrations of the key metabolites of hepatic energy metabolism, adenosine triphosphate (ATP) and inorganic phosphate (Pi), can be altered in metabolic disorders such as diabetes mellitus. 31Phosphorus (31P)‐magnetic resonance spectroscopy (MRS) is used to noninvasively measure hepatic metabolites, but measuring their absolute molar concentrations remains challenging. This study employed a 31P‐MRS method based on the phantom replacement technique for quantifying hepatic 31P‐metabolites on a 3‐T clinical scanner. Two surface coils with different size and geometry were used to check for consistency in terms of repeatability and reproducibility and absolute concentrations of metabolites. Day‐to‐day (n = 8) and intra‐day (n = 6) reproducibility was tested in healthy volunteers. In the day‐to‐day study, mean absolute concentrations of γ‐ATP and Pi were 2.32 ± 0.24 and 1.73 ± 0.26 mM (coefficient of variation [CV]: 7.3% and 8.8%) for the single loop, and 2.32 ± 0.42 and 1.73 ± 0.27 mM (CVs 6.7% and 10.6%) for the quadrature coil, respectively. The intra‐day study reproducibility using the quadrature coil yielded CVs of 4.7% and 6.8% for γ‐ATP and Pi without repositioning, and 6.3% and 7.1% with full repositioning of the volunteer. The results of the day‐to‐day data did not differ between coils and visits. Both coils robustly yielded similar results for absolute concentrations of hepatic 31P‐metabolites. The current method, applied with two different surface coils, can be readily utilized in long‐term and interventional studies. In comparison with the single loop coil, the quadrature coil also allows measurements at a greater distance between the coil and liver, which is relevant for studying people with obesity.

Funder

Bundesministerium für Bildung und Forschung

Bundesministerium für Gesundheit

European Research Council

Schmutzler Stiftung

Deutsche Forschungsgemeinschaft

HORIZON EUROPE Framework Programme

Ministerium für Kultur und Wissenschaft des Landes Nordrhein-Westfalen

Publisher

Wiley

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3