Development of a three‐dimensional scintillation detector for pencil beam verification in proton therapy patient‐specific quality assurance

Author:

Frelin Anne‐Marie1,Daviau Gautier12,Bui My Hoang Hoa1,Fontbonne Cathy3,Fontbonne Jean‐Marc3,Lebhertz Dorothée34,Mainguy Erwan1,Moignier Cyril34,Thariat Juliette34,Vela Anthony34

Affiliation:

1. Grand accélérateur National dʼIons Lourds (GANIL), CEA/DRF‐CNRS/IN2P3 Caen France

2. Normandie University, UNICAEN Caen France

3. Université de Caen Normandie ENSICAEN CNRS/IN2P3 Caen France

4. Medical Physics Department CLCC François Baclesse Caen France

Abstract

AbstractBackgroundPencil Beam Scanning proton therapy has many advantages from a therapeutic point of view, but raises technical constraints in terms of treatment verification. The treatment relies on a large number of planned pencil beams (PB) (up to thousands), whose delivery is divided in several low‐intensity pulses delivered a high frequency (1 kHz in this study).PurposeThe purpose of this study was to develop a three‐dimensional quality assurance system allowing to verify all the PBs’ characteristics (position, energy, intensity in terms of delivered monitor unit—MU) of patient treatment plans on a pulse‐by‐pulse or a PB‐by‐PB basis.MethodsA system named SCICOPRO has been developed. It is based on a 10 × 10 × 10 cm3 scintillator cube and a fast camera, synchronized with beam delivery, recording two views (direct and using a mirror) of the scintillation distribution generated by the pulses. A specific calibration and analysis process allowed to extract the characteristics of all the pulses delivered during the treatment, and consequently of all the PBs. The system uncertainties, defined here as average value + standard deviation, were characterized with a customized irradiation plan at different PB intensities (0.02, 0.1, and 1 MU) and with two patient's treatment plans of three beams each. The system's ability to detect potential treatment delivery problems, such as positioning errors of the treatment table in this work (1° rotations and a 2 mm translation), was assessed by calculating the confidence intervals (CI) for the different characteristics and evaluating the proportion of PBs within these intervals.ResultsThe performances of SCICOPRO were evaluated on a pulse‐by‐pulse basis. They showed a very good signal‐to‐noise ratio for all the pulse intensities (between 2 × 10−3 MU and 150 × 10−3 MU) allowing uncertainties smaller than 580 µm for the position, 180 keV for the energy and 3% for the intensity on patients treatment plans. The position and energy uncertainties were found to be little dependent from the pulse intensities whereas the intensity uncertainty depends on the pulses number and intensity distribution. Finally, treatment plans evaluations showed that 98% of the PBs were within the CIs with a nominal positioning against 83% or less with the table positioning errors, thus proving the ability of SCICOPRO to detect this kind of errors.ConclusionThe high acquisition rate and the very high sensitivity of the system developed in this work allowed to record pulses of intensities as low as 2 × 10−3 MU. SCICOPRO was thus able to measure all the characteristics of the spots of a treatment (position, energy, intensity) in a single measurement, making it possible to verify their compliance with the treatment plan. SCICOPRO thus proved to be a fast and accurate tool that would be useful for patient‐specific quality assurance (PSQA) on a pulse‐by‐pulse or PB‐by‐PB verification basis.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3