Factors affecting performance of fetal blood T2 measurements for noninvasive estimation of oxygen saturation

Author:

Hedström Erik1234ORCID,Piek Marjolein13ORCID,Bidhult‐Johansson Sebastian1,Ryd Daniel13,Testud Frederik5,Töger Johannes1ORCID,Aletras Anthony H.16

Affiliation:

1. Clinical Physiology, Department of Clinical Sciences Lund Lund University Lund Sweden

2. Diagnostic Radiology, Department of Clinical Sciences Lund Lund University Lund Sweden

3. Department of Clinical Physiology and Nuclear Medicine Skåne University Hospital Lund Sweden

4. Department of Radiology Skåne University Hospital Lund Sweden

5. Siemens Healthcare AB Malmö Sweden

6. Laboratory of Computing, Medical Informatics and Biomedical‐Imaging Technologies, School of Medicine Aristotle University of Thessaloniki Thessaloniki Greece

Abstract

AbstractPurposeTo ultimately make accurate and precise fetal noninvasive oxygen saturation (sO2) measurements by T2‐prepared bSSFP more widely available by systematically assessing error sources in order to potentially reduce perinatal mortality in cardiovascular malformations and fetal growth restriction.MethodsT2‐prepared bSSFP data were acquired in phantoms; in flowing blood in adults in the superior sagittal sinus, ascending and descending aorta, and main pulmonary artery; and in the fetal descending aorta and umbilical vein. T2 was assessed in relation to T2 two‐ or three‐parameter curve‐fitting techniques, SSFP readout, refocusing time delay (τ), constant and pulsatile blood flow, and impact of T1 recovery. Further, fetal T2 and sO2 variability were quantified in the descending aorta and umbilical vein in healthy fetuses and fetuses with cardiovascular malformation (gestational weeks 32–38).ResultsIn phantoms, three‐parameter fitting was accurate irrespective of phase FOV (<4 ms; i.e., <2%), and T2 was overestimated (up to 23 ms/10%; p = 0.001) beyond ±30 Hz off‐resonance. In the adult aorta, T2 was underestimated during higher blood flow velocities and pulsatility for τ = 16 ms (−41 ms/−17%; p = 0.008). In fetuses, two‐parameter fitting overestimated T2 compared with three‐parameter fitting (+33 ms/+18%; p = 0.03). T2 variability was 18 ms/15% in the fetal descending aorta and 28 ms/14% in the umbilical vein. The resulting estimated sO2 variability was ∼10% (15% of sO2 value) in the fetal descending aorta.ConclusionsErrors due to T2‐fitting techniques, off‐resonance, flow velocity, and insufficient T1 recovery between image acquisitions could be mitigated by using three‐parameter fitting with included saturation‐prepared images approximating infinite T2‐preparation time, adequate shimming covering the fetus and placenta, and by modifying acquisition parameters. Variability in fetal blood T2 and sO2, however, indicate that it is currently not feasible to use these methods for prediction of disease.

Funder

Hjärt-Lungfonden

Region Skåne

Skånes universitetssjukhus

Vetenskapsrådet

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3