Energy‐based scatter estimation in clinical PET

Author:

Hamill James J.1,Cabello Jorge1,Surti Suleman2,Karp Joel S.2

Affiliation:

1. Siemens Medical Solutions Knoxville Tennessee USA

2. Department of Radiology Perelman School of Medicine University of Pennsylvania Philadelphia Pennsylvania USA

Abstract

AbstractBackgroundScatter correction (SC) is essential in PET for accurate quantitative imaging. The state‐of‐the‐art SC method is single‐scatter simulation (SSS). Although this method is usually robust and accurate, it can fail in some situations, for example when there is motion between the CT and PET scans in PET/CT. Therefore, it is of interest to consider other SC methods.PurposeIn this work, an energy‐based scatter estimation (EBS) method is described in detail, tested in phantoms and patients, and compared to SSS.MethodsThis version of EBS was developed for list‐mode data from Biograph Vision‐600 PET/CT scanner. EBS is based on digitized 2D energy histograms in each bin of a coarsely sampled PET sinogram, either with or without time of flight (TOF). The histograms are modeled as a noisy realization of a linear combination of nine basis functions whose parameters were derived from a measurement of the 511‐keV photopeak spectrum as well as Monte‐Carlo simulations of the scattering process. EBS uses an iterative expectation maximization approach to determine the coefficients in the linear combination, and from this estimates the scatter. The investigation was restricted to 18F‐based PET data in which the acquired number of counts was similar to the levels seen in oncological whole‐body PET/CT scans. To evaluate the performance, phantom scans were used that involved the NEMA NU2‐2018 protocol, a slab phantom, an NU 2‐1994 phantom, a cardiac phantom in an anthropomorphic chest phantom, and a uniformly‐filled torso phantom with a bladder phantom slightly outside the axial field of view. Contrast recovery (CR) and other parameters were evaluated in images reconstructed with SSS and EBS. Furthermore, FDG PET scans of seven lung cancer patients were used in the evaluation. Standardized uptake values (SUV) based on SSS and EBS were compared in 27 lesions.ResultsEBS and SSS images were visually similar in all cases except the torso + bladder phantom, where the EBS was much closer to the expected uniform image. The NU2‐2018 analysis indicated a 2% scatter residual in EBS images compared to 3% with SSS, and 10% higher background variability, which is a surrogate for image noise. The cardiac phantom scan showed that CR was 98.2% with EBS and 99.6% with SSS, and that the SSS sinogram had values greater than the net‐true emission sinogram, indicating a slight overcorrection in the case of SSS. In the lesion SUV comparison in patient scans, EBS correlated strongly (R2 = 0.9973) with SSS, and SUV based on EBS were systematically 0.1 SUV lower. In the case of the torso + bladder phantom portion, the SSS image of the torso + bladder phantom was 299% times hotter than expected in one area, due to scatter estimation error, compared to 16% colder with EBS.ConclusionsIn evaluating clinically relevant parameters such as SUV in focal lesions, EBS and SSS give almost the same results. In phantoms, some scatter figures of merit were slightly improved by use of EBS, though an image variability figure of merit was slightly degraded. In typical oncological whole‐body PET/CT, EBS may be a suitable replacement for SSS, especially when SSS fails due to technical problems during the scan.

Funder

National Institutes of Health

Publisher

Wiley

Subject

General Medicine

Reference25 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3