Ultrasound‐guided breast biopsy using an adapted automated cone‐based ultrasound scanner: a feasibility study

Author:

Nikolaev Anton V.1,de Jong Leon1,Zamecnik Patrik1,Groenhuis Vincent2,Siepel Françoise J.2,Stramigioli Stefano2,Hansen Hendrik H.G.1,de Korte Chris L.13

Affiliation:

1. Medical Ultrasound Imaging Center (MUSIC), Department of Medical Imaging/Radiology Radboud University Medical Center Nijmegen The Netherlands

2. Robotics and Mechatronics University of Twente Enschede The Netherlands

3. Physics of Fluids Group, TechMed Center University of Twente Enschede The Netherlands

Abstract

AbstractBackgroundAmong available breast biopsy techniques, ultrasound (US)‐guided biopsy is preferable because it is relatively inexpensive and provides live imaging feedback. The availability of magnetic resonance imaging (MRI)‐3D US image fusion would facilitate US‐guided biopsy even for US occult lesions to reduce the need for expensive and time‐consuming MRI‐guided biopsy. In this paper, we propose a novel Automated Cone‐based Breast Ultrasound Scanning and Biopsy System (ACBUS‐BS) to scan and biopsy breasts of women in prone position. It is based on a previously developed system, called ACBUS, that facilitates MRI‐3D US image fusion imaging of the breast employing a conical container filled with coupling medium.PurposeThe purpose of this study was to introduce the ABCUS‐BS system and demonstrate its feasibility for biopsy of US occult lesions.MethodThe biopsy procedure with the ACBUS‐BS comprises four steps: target localization, positioning, preparation, and biopsy. The biopsy outcome can be impacted by 5 types of errors: due to lesion segmentation, MRI‐3D US registration, navigation, lesion tracking during repositioning, and US inaccuracy (due to sound speed difference between the sample and the one used for image reconstruction).For the quantification, we use a soft custom‐made polyvinyl alcohol phantom (PVA) containing eight lesions (three US‐occult and five US‐visible lesions of 10 mm in diameter) and a commercial breast mimicking phantom with a median stiffness of 7.6 and 28 kPa, respectively. Errors of all types were quantified using the custom‐made phantom. The error due to lesion tracking was also quantified with the commercial phantom.Finally, the technology was validated by biopsying the custom‐made phantom and comparing the size of the biopsied material to the original lesion size. The average size of the 10‐mm‐sized lesions in the biopsy specimen was 7.00 ± 0.92 mm (6.33 ± 1.16 mm for US occult lesions, and 7.40 ± 0.55 mm for US‐visible lesions).ResultsFor the PVA phantom, the errors due to registration, navigation, lesion tracking during repositioning, and US inaccuracy were 1.33, 0.30, 2.12, and 0.55 mm. The total error was 4.01 mm. For the commercial phantom, the error due to lesion tracking was estimated at 1.10 mm, and the total error was 4.11 mm. Given these results, the system is expected to successfully biopsy lesions larger than 8.22 mm in diameter. Patient studies will have to be carried out to confirm this in vivo.ConclusionThe ACBUS‐BS facilitates US‐guided biopsy of lesions detected in pre‐MRI and therefore might offer a low‐cost alternative to MRI‐guided biopsy. We demonstrated the feasibility of the approach by successfully taking biopsies of five US‐visible and three US‐occult lesions embedded in a soft breast‐shaped phantom.

Publisher

Wiley

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3