Validation of an EGSnrc‐based Monte Carlo model for a complex 2D‐array for technical QA measurements of a linear accelerator

Author:

Czarnecki Damian1,Zink Klemens123,Alissa Mohamad12,Flatten Veronika123,Espelage Teresa4,Schoenfeld Andreas A.5

Affiliation:

1. Institute of Medical Physics and Radiation Protection University of Applied Sciences Giessen (THM) Giessen Germany

2. Department of Radiotherapy and Radiation Oncology University Medical Center Giessen and Marburg Marburg Germany

3. Marburg Ionbeam Therapy Center (MIT) Marburg Germany

4. Radiologie Vechta Vechta Germany

5. Sun Nuclear Corporation Melbourne Florida USA

Abstract

AbstractPurposeMulti‐axis ionization chamber arrays can be used for quality assurance (QA) and measurement of linear accelerator (linac) specific data. In this work, the ability of the IC Profiler (Sun Nuclear Corp., Melbourne, Florida) detector array to measure the photon beam quality specifier %dd(10) and TPR20, 10 was investigated. To investigate the method for beam energy QA using a two‐dimensional detector array, a Monte Carlo‐based model of the detector array was developed and validated.MethodsA Monte Carlo‐based model of the IC Profiler detector array with Quad Wedge accessories was developed in detail from drawings provided by the manufacturer using the egs++ class library from the EGSnrc code system. Monte Carlo simulations were used to calculate the absorbed dose in the 251 ionization chambers of the IC profiler in the 6 MV Elekta Precise radiation field. To validate the results from the Monte Caro simulations, measurements were performed on clinical 6 MV linacs. To vary the photon beam quality of the Elekta 6 MV linac, the current of the bending magnet was varied. Furthermore, the area ratio was calculated from IC Profiler measurements with the Quad Wedge accessories.ResultsMeasurements as well as Monte Carlo simulations confirmed the linear relationship between the area ratio and the investigated photon beam quality specifier %dd(10) and TPR20, 10 for the investigated radiation source. Furthermore, the Monte Carlo‐simulated data were within the 95% confidence interval of the linear fit to the measured data. This enabled the Monte Carlo‐based IC Profiler model to be used for further investigations. The values were calculated for various electron beam sizes and the angle of incidence on the target of the linac.ConclusionsA Monte‐Carlo‐based model of the detector array was developed, which could successfully reproduce measurements, demonstrating that even very complex geometries can be modeled in EGSnrc. Moreover, the study showed that the validated Monte Carlo model has the potential to investigate variations in beam parameters and their effects on AR ratios and %dd(10) that may not be investigated experimentally. While these findings may help users gain a deeper understanding of the QA method, the Monte Carlo model enables other complex investigations, such as the simulation of measurements in the presence of magnetic fields, or the simulation of measurements on novel treatment delivery techniques and devices.

Publisher

Wiley

Subject

General Medicine

Reference18 articles.

1. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

2. AndreoP BurnsDT HohlfeldK et al.Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. Technical Reports Series No. 398.International Atomic Energy Agency;2006.

3. Task Group 142 report: Quality assurance of medical acceleratorsa)

4. Beam energy metrics for the acceptance and quality assurance of Halcyon linear accelerator

5. Clinical implementation of photon beam flatness measurements to verify beam quality