Validation of cardiac index measured by four‐dimensional cardiac magnetic resonance flow against the invasively measured cardiac index in patients with pulmonary hypertension

Author:

Romeih Soha12ORCID,Elkafrawy Fatma13ORCID,Shaaban Mahmoud12ORCID,Hassan Ahmed45,Samuel Irini45,Gergis Micheal5ORCID,El mozy Wesam16,Yacoub Magdi17

Affiliation:

1. Radiology Department Aswan Heart Centre Aswan Egypt

2. Cardiology Department Tanta University Tanta Egypt

3. Radiology Department Alexandria University Alexandria Egypt

4. Cardiology Department Cairo University Giza Egypt

5. Cardiology Department Aswan Heart Centre Aswan Egypt

6. Radiology Department Cairo University Giza Egypt

7. Imperial Collage of London London UK

Abstract

AbstractThe purpose of this study was to validate cardiac index (CI) measured noninvasively by four‐dimensional (4D) and two‐dimensional (2D) cardiovascular magnetic resonance (CMR) flows against the invasively measured CI by right heart catheterization (RHC) in patients with pulmonary hypertension (PH). Thirty patients with PH (mean age: 32 ± 10 years) were included. 4D and 2D flow measured CI within 24 h from RHC measured CI invasively. Qualitative analysis of 4D pulmonary flow (vortex presence and eccentricity of flow) was performed.All patients had helical right‐sided flow with vortex formation; the mean vortex diameter was 29 ± 7 mm, occupying 69% of the main pulmonary artery (MPA) lumen. MPA was dilated (42 ± 9 mm). Mean CI measured by 4D flow CMR was closer to mean CI measured invasively (indirect Fick method CI = 2.1 ± 0.8 L/min/m2 vs. PA 4D flow = 2.3 ± 0.7 L/min/m2 “bias 0.22 ± 0.25 L/min, p = 0.001,” and Ao 4D flow = 2.3 ± 0.7 L/min/m2 “bias 0.2 ± 0.28 L/min, p = 0.001”), while 2D flow had a higher mean CI (PA 2D flow = 2.5 ± 0.7 L/min/m2 “bias 0.45 ± 0.7 L/min, p = 0.001” and Ao 2D flow = 2.5 ± 0.8 L/min/m2 “bias 0.45 ± 0.67 L/min, p = 0.001”). The correlation coefficients among the different comparisons of CI showed: a low correlation between the 2D flow‐indirect Fick method (Ao r2 = 0.37, PA r2 = 0.32) and a high correlation between the 4D flow‐indirect Fick method (Ao r2 = 0.86, PA r2 = 0.89).There is an excellent agreement between CI measured by 4D flow and CI measured invasively. 4D flow, a noninvasive imaging technique, could accurately measure CI better than the conventional 2D flow in patients with PH.

Publisher

Wiley

Subject

Pulmonary and Respiratory Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献