Affiliation:
1. Obesity and Diabetes Clinical Research Section, Phoenix Epidemiology and Clinical Research Branch National Institute of Diabetes and Digestive and Kidney Diseases Phoenix Arizona USA
2. Department of Information Engineering University of Pisa Pisa Italy
3. MEI Research, Ltd. Edina Minnesota USA
Abstract
AbstractObjectiveThe aim of this study was to test proportional‐integral‐derivative (PID) control of air inflow rate in a whole‐room indirect calorimeter to improve accuracy in measuring oxygen (O2) consumption () and carbon dioxide (CO2) production ().MethodsA precision gas blender infused nitrogen (N2) and CO2 into the calorimeter over 24 hours based on static and dynamic infusion profiles mimicking and patterns during resting and non‐resting conditions. Constant (60 L/min) versus time‐variant flow set by a PID controller based on the CO2 concentration was compared based on errors between measured versus expected values for respiratory exchange ratio, and metabolic rate.ResultsCompared with constant inflow, the PID controller allowed both a faster rise time and long‐term maintenance of a stable CO2 concentration inside the calorimeter, resulting in more accurate estimates (mean hourly error, PID: −0.9%, 60 L/min = −2.3%, p < 0.05) during static infusions. During dynamic infusions mimicking exercise sessions, the PID controller achieved smaller errors for (mean: −0.6% vs. −2.7%, p = 0.02) and respiratory exchange ratio (mean: 0.5% vs. −3.1%, p = 0.02) compared with constant inflow conditions, with similar (p = 0.97) and metabolic rate (p = 0.76) errors.ConclusionsPID control in a whole‐room indirect calorimeter system leads to more accurate measurements of substrate oxidation during dynamic metabolic studies.
Funder
National Institute of Diabetes and Digestive and Kidney Diseases
National Institutes of Health
Subject
Nutrition and Dietetics,Endocrinology,Endocrinology, Diabetes and Metabolism,Medicine (miscellaneous)
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献