Electrospray mass spectrometry method to prevent the reduction of hexavalent to pentavalent chromium

Author:

Muto Miu1,Osuka Yuri1,Ii Kyota1,Ueda Richi1,Shintani Ayumi1,Takanashi Midori1,Terada Itsumi1,Oda Kyohei1,Tsuchiya Kotaro1,Sahoo Yu Vin2,Takahashi Kazuya2ORCID,Tanaka Miho1ORCID

Affiliation:

1. Graduate School of Marine Science and Technology Tokyo University of Marine Science and Technology Minato‐ku Tokyo Japan

2. RIKEN Wako‐shi Saitama Japan

Abstract

RationaleElectrospray mass spectrometry (ESI‐MS) is one of the most effective methods for assessing the state of metals in solution. For ions with a redox potential close to ~0.55 V, such as Cr6+, reduction of the metal in solution occurs in the ESI‐MS system. In our studies, it was observed that [HCrO4] undergoes reduction, resulting in the formation of [CrO3]. The precise mechanism remains ambiguous. The reduction of hexavalent chromium to pentavalent chromium is supported by Frost diagrams, reinforcing our confidence in the validity of the ESI‐MS measurement method. The reduction mechanism in ESI‐MS was clarified, and a system was devised to eliminate electron donation during the reduction of Cr6+ in solution.MethodsTo determine the state of Cr6+ by ESI‐MS, CrO3 in solid form was dissolved in ultrapure water to prepare a solution of 500 × 10−6 mol/L (μM) concentration. The pH was adjusted to 4.0, 5.3, 6.3, 8.2 and 9.1 and subsequently measured. CrO3 solutions with various concentrations of 10, 100 and 500 μM were prepared and adjusted to a pH of ~7 using tetramethylammonium hydroxide to measure Cr6+ under different conditions.ResultsCr6+ in solution was soluble and existed as an oxoacid with a negative charge independent of pH. Cr6+ was stable over a wide pH range at various concentrations. The ESI‐MS method determined the negative ion [HCrO4] as the stable ion, but [CrO3] was also present as a byproduct. Therefore, we were interested in the presence of other species, such as [CrO3], which could have formed owing to the reduction of Cr6+.ConclusionsIn ESI‐MS system, it undergoes reduction to form [CrO3]. The high flow rate of ultrapure water in pump insulated the acceptance of electrons by Cr6+ preventing its reduction. Further in‐depth ESI‐MS studies could explain the complex formation and behavior of Cr6+ in aqueous solution.

Funder

Toyo Suisan Foundation

Salt Science Research Foundation

Publisher

Wiley

Subject

Organic Chemistry,Spectroscopy,Analytical Chemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3