Affiliation:
1. MRI System Engineering Team MICROPORT CRM Clamart France
Abstract
PurposeThe aim of this paper is to study the impact of pulse generator small impedance variations on the overall radiofrequency (RF)‐induced heating of a cardiac rhythm management (CRM) lead in a 1.5 T MR system.MethodsAn analysis was conducted on two CRM pacing leads with different structural designs. The leads were connected to three CRM devices belonging to different technological platforms: two pacemakers and a cardiac resynchronization therapy defibrillator (CRT‐D). All pulse generators had a small (< ) RF input impedance compared to the overall system (pulse generator and lead) impedance. In addition, two Dummy load devices with an input impedance that is not negligible compared to the lead impedance were also studied. The implantable systems were studied in terms of RF deposited power at the lead electrodes, which is proportional to the RF‐induced heating.ResultsResults show that both leads behave very similarly in terms of RF deposited power for the same in‐vivo tangential E‐field excitation when mounting any of the pulse generators; reported differences are well below the combined uncertainty . On the other hand, both Dummy load devices resulted in a RF deposited power which differs from the other pulse generators by values larger than the combined uncertainty.ConclusionThe results of our experimental analysis show that different pulse generators would not change the RF‐induced heating behavior of a given lead, provided that the pulse generator impedances are small (in our study less than 5%) compared to the overall system impedance.
Subject
Radiology, Nuclear Medicine and imaging
Reference21 articles.
1. Determining the risks of clinically indicated nonthoracic magnetic resonance imaging at 1.5 T for patients with pacemakers and implantable cardioverter-defibrillators: Rationale and design of the MagnaSafe Registry
2. ISO/TS 10974:2018.Assessment of the safety of magnetic resonance imaging for patients with an active implantable medical device. Technical report International Organization for Standardization;2018.
3. PrutchiD MeyersJ ShehadaR.RF impedance of MR‐conditional pacemaker leads when connected to implantable pulse generators from different MR‐conditional systems. Paper presented at: Proceedings of the 2021 Annual Meeting of ISMRM and SMRT Online;2021:2281.
4. On the Relationship Between Impedances of Active Implantable Medical Devices and Device Safety Under MRI RF Emission