Denoising magnetic resonance spectroscopy (MRS) data using stacked autoencoder for improving signal‐to‐noise ratio and speed of MRS

Author:

Wang Jing1,Ji Bing2,Lei Yang1,Liu Tian3,Mao Hui2,Yang Xiaofeng1

Affiliation:

1. Department of Radiation Oncology and Winship Cancer Institute Emory University Atlanta USA

2. Department of Radiology and Imaging Science and Winship Cancer Institute Emory University Atlanta USA

3. Department of Radiation Oncology Icahn School of Medicine at Mount Sinai New York USA

Abstract

AbstractBackgroundWhile magnetic resonance imaging (MRI) provides high resolution anatomical images with sharp soft tissue contrast, magnetic resonance spectroscopy (MRS) enables non‐invasive detection and measurement of biochemicals and metabolites. However, MRS has low signal‐to‐noise ratio (SNR) when concentrations of metabolites are in the range of millimolar. Standard approach of using a high number of signal averaging (NSA) to achieve sufficient SNR comes at the cost of a long acquisition time.PurposeWe propose to use deep‐learning approaches to denoise MRS data without increasing NSA. This method has potential to reduce the acquisition time as well as improve SNR and quality of spectra, which could enhance the diagnostic value and broaden the clinical applications of MRS.MethodsThe study was conducted using data collected from the brain spectroscopy phantom and human subjects. We utilized a stack auto‐encoder (SAE) network to train deep learning models for denoising low NSA data (NSA = 1, 2, 4, 8, and 16) randomly truncated from high SNR data collected with high NSA (NSA = 192), which were also used to obtain the ground truth. We applied both self‐supervised and fully‐supervised training approaches and compared their performance of denoising low NSA data based on improvement in SNR. To prevent overfitting, the SAE network was trained in a patch‐based manner. We then tested the denoising methods on noise‐containing data collected from the phantom and human subjects, including data from brain tumor patients. We evaluated their performance by comparing the SNR levels and mean squared errors (MSEs) calculated for the whole spectra against high SNR “ground truth”, as well as the value of chemical shift of N‐acetyl‐aspartate (NAA) before and after denoising.ResultsWith the SAE model, the SNR of low NSA data (NSA = 1) obtained from the phantom increased by 28.5% and the MSE decreased by 42.9%. For low NSA data of the human parietal and temporal lobes, the SNR increased by 32.9% and the MSE decreased by 63.1%. In all cases, the chemical shift of NAA in the denoised spectra closely matched with the high SNR spectra without significant distortion to the spectra after denoising. Furthermore, the denoising performance of the SAE model was more effective in denoising spectra with higher noise levels.ConclusionsThe reported SAE denoising method is a model‐free approach to enhance the SNR of MRS data collected with low NSA. With the denoising capability, it is possible to acquire MRS data with a few NSA, shortening the scan time while maintaining adequate spectroscopic information for detecting and quantifying the metabolites of interest. This approach has the potential to improve the efficiency and effectiveness of clinical MRS data acquisition by reducing the scan time and increasing the quality of spectroscopic data.

Funder

National Institutes of Health

Publisher

Wiley

Subject

General Medicine

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Spectro-ViT: A vision transformer model for GABA-edited MEGA-PRESS reconstruction using spectrograms;Magnetic Resonance Imaging;2024-11

2. Results of the 2023 ISBI challenge to reduce GABA-edited MRS acquisition time;Magnetic Resonance Materials in Physics, Biology and Medicine;2024-04-13

3. Machine Learning for Image Denoising: A Review;Lecture Notes in Electrical Engineering;2024

4. An Investigation of Different Deep Learning Pipelines for GABA-Edited MRS Reconstruction;Machine Learning in Medical Imaging;2023-10-15

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3