Faster and lower dose imaging: evaluating adaptive, constant gantry velocity and angular separation in fast low‐dose 4D cone beam CT imaging

Author:

Lau Benjamin K. F.1,Dillon Owen1,Vinod Shalini K.23,O'Brien Ricky T.14,Reynolds Tess1

Affiliation:

1. Faculty of Medicine and Health Image X Institute University of Sydney Sydney NSW Australia

2. Liverpool & Macarthur Cancer Therapy Centres Liverpool Hospital Liverpool New South Wales Australia

3. South Western Sydney Clinical School The University of New South Wales & Ingham Institute for Applied Medical Research Liverpool New South Wales Australia

4. Medical Radiations School of Health and Biomedical Sciences RMIT University Bundoora Victoria Australia

Abstract

AbstractBackgroundThe adoption of four‐dimensional cone beam computed tomography (4DCBCT) for image‐guided lung cancer radiotherapy is increasing, especially for hypofractionated treatments. However, the drawbacks of 4DCBCT include long scan times (∼240 s), inconsistent image quality, higher imaging dose than necessary, and streaking artifacts. With the emergence of linear accelerators that can acquire 4DCBCT scans in a short period of time (9.2 s) there is a need to examine the impact that these very fast gantry rotations have on 4DCBCT image quality.PurposeThis study investigates the impact of gantry velocity and angular separation between x‐ray projections on image quality and its implication for fast low‐dose 4DCBCT with emerging systems, such as the Varian Halcyon that provide fast gantry rotation and imaging. Large and uneven angular separation between x‐ray projections is known to reduce 4DCBCT image quality through increased streaking artifacts. However, it is not known when angular separation starts degrading image quality. The study assesses the impact of constant and adaptive gantry velocity and determines the level when angular gaps impair image quality using state‐of‐the‐art reconstruction methods.MethodsThis study considers fast low‐dose 4DCBCT acquisitions (60–80 s, 200‐projection scans). To assess the impact of adaptive gantry rotations, the angular position of x‐ray projections from adaptive 4DCBCT acquisitions from a 30‐patient clinical trial were analyzed (referred to as patient angular gaps). To assess the impact of angular gaps, variable and static angular gaps (20°, 30°, 40°) were introduced into evenly separated 200 projections (ideal angular separation). To simulate fast gantry rotations, which are on emerging linacs, constant gantry velocity acquisitions (9.2 s, 60 s, 120 s, 240 s) were simulated by sampling x‐ray projections at constant intervals using the patient breathing traces from the ADAPT clinical trial (ACTRN12618001440213). The 4D Extended Cardiac‐Torso (XCAT) digital phantom was used to simulate projections to remove patient‐specific image quality variables.Image reconstruction was performed using Feldkamp‐Davis‐Kress (FDK), McKinnon‐Bates (MKB), and Motion‐Compensated‐MKB (MCMKB) algorithms. Image quality was assessed using Structural Similarity‐Index‐Measure (SSIM), Contrast‐to‐Noise‐Ratio (CNR), Signal‐to‐Noise‐Ratio (SNR), Tissue‐Interface‐Width‐Diaphragm (TIW‐D), and Tissue‐Interface‐Width‐Tumor (TIW‐T).ResultsPatient angular gaps and variable angular gap reconstructions produced similar results to ideal angular separation reconstructions, while static angular gap reconstructions produced lower image quality metrics. For MCMKB‐reconstructions, average patient angular gaps produced SSIM‐0.98, CNR‐13.6, SNR‐34.8, TIW‐D‐1.5 mm, and TIW‐T‐2.0 mm, static angular gap 40° produced SSIM‐0.92, CNR‐6.8, SNR‐6.7, TIW‐D‐5.7 mm, and TIW‐T‐5.9 mm and ideal produced SSIM‐1.00, CNR‐13.6, SNR‐34.8, TIW‐D‐1.5 mm, and TIW‐T‐2.0 mm. All constant gantry velocity reconstructions produced lower image quality metrics than ideal angular separation reconstructions regardless of the acquisition time. Motion compensated reconstruction (MCMKB) produced the highest contrast images with low streaking artifacts.ConclusionVery fast 4DCBCT scans can be acquired provided that the entire scan range is adaptively sampled, and motion‐compensated reconstruction is performed. Importantly, the angular separation between x‐ray projections within each individual respiratory bin had minimal effect on the image quality of fast low‐dose 4DCBCT imaging. The results will assist the development of future 4DCBCT acquisition protocols that can now be achieved in very short time frames with emerging linear accelerators.

Funder

National Health and Medical Research Council

Publisher

Wiley

Subject

General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3