Proton dosimetry in a magnetic field: Measurement and calculation of magnetic field correction factors for a plane‐parallel ionization chamber

Author:

Gebauer Benjamin12,Baumann Kilian‐Simon345,Fuchs Hermann67,Georg Dietmar67,Oborn Brad M.289,Looe Hui‐Khee10,Lühr Armin11

Affiliation:

1. OncoRay National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden Helmholtz‐Zentrum Dresden‐Rossendorf Dresden Germany

2. Institute of Radiooncology‐OncoRay Helmholtz‐Zentrum Dresden‐Rossendorf Dresden Germany

3. Department of Radiotherapy and Radiooncology University Medical Center Giessen‐Marburg Marburg Germany

4. University of Applied Sciences Institute of Medical Physics and Radiation Protection Giessen Germany

5. Ion‐Beam Therapy Center Marburg Germany

6. Department of Radiation Oncology Medical University of Vienna Wien Austria

7. MedAustron Iontherapy centre Wiener Neustadt Austria

8. Centre for Medical Radiation Physics University of Wollongong Wollongong New South Wales Australia

9. Illawarra Cancer Care Centre Wollongong New South Wales Australia

10. Department for Radiotherapy and Radiooncology Pius Hospital, Medical Campus Carl von Ossietzky University Oldenburg Germany

11. Department of Physics TU Dortmund University Dortmund Germany

Abstract

AbstractBackgroundThe combination of magnetic resonance imaging and proton therapy offers the potential to improve cancer treatment. The magnetic field (MF)‐dependent change in the dosage of ionization chambers in magnetic resonance imaging‐integrated proton therapy (MRiPT) is considered by the correction factor , which needs to be determined experimentally or computed via Monte Carlo (MC) simulations.PurposeIn this study, was both measured and simulated with high accuracy for a plane‐parallel ionization chamber at different clinical relevant proton energies and MF strengths.Material and methodsThe dose‐response of the Advanced Markus chamber (TM34045, PTW, Freiburg, Germany) irradiated with homogeneous 10× 10 cm2 quasi mono‐energetic fields, using 103.3, 128.4, 153.1, 223.1, and 252.7 MeV proton beams was measured in a water phantom placed in the MF of an electromagnet with MF strengths of 0.32, 0.5, and 1 T. The detector was positioned at a depth of 2 g/cm2 in water, with chamber electrodes parallel to the MF lines and perpendicular to the proton beam incidence direction. The measurements were compared with TOPAS MC simulations utilizing COMSOL‐calculated 0.32, 0.5, and 1 T MF maps of the electromagnet. was calculated for the measurements for all energies and MF strengths based on the equation: , where and MQ were the temperature and air‐pressure corrected detector readings with and without the MF, respectively. MC‐based correction factors were determined as , where and Ddet were the doses deposited in the air cavity of the ionization chamber model with and without the MF, respectively. Furthermore, MF effects on the chamber dosimetry are studied using MC simulations, examining the impact on the absorbed dose‐to‐water () and the shift in depth of the Bragg peak.ResultsThe detector showed a reduced dose‐response for all measured energies and MF strengths, resulting in experimentally determined values larger than unity. For all energies and MF strengths examined, ranged between 1.0065 and 1.0205. The dependence on the energy and the MF strength was found to be non‐linear with a maximum at 1 T and 252.7 MeV. The MC simulated values agreed with the experimentally determined correction factors within their standard deviations with a maximum difference of 0.6%. The MC calculated impact on was smaller 0.2 %.ConclusionFor the first time, measurements and simulations were compared for proton dosimetry within MFs using an Advanced Markus chamber. Good agreement of was found between experimentally determined and MC calculated values. The performed benchmarking of the MC code allows for calculating for various ionization chamber models, MF strengths and proton energies to generate the data needed for a proton dosimetry protocol within MFs and is, therefore, a step towards MRiPT.

Funder

European Metrology Programme for Innovation and Research

Publisher

Wiley

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3