Affiliation:
1. Department of Chemistry GITAM School of Science Hyderabad India
Abstract
AbstractGlipizide is an antidiabetic drug used for the treatment of type 2 diabetes. A simple, reliable and robust reverse‐phase liquid chromatographic method (RP‐HPLC) was developed and validated as per International Conference on Harmonization Q2(R1) for estimating the impurities of glipizide in pharmaceutical formulations. The chromatographic separation was carried out on a Phenomenex Luna C18 (2), 250 × 4.6 mm, 5 μm with a binary solvent delivery system [MP‐A, a homogenous mixture of water and acetonitrile in a ratio of 90:10 (v/v) and 1 ml of orthophosphoric acid; and MP‐B, a homogenous mixture of water and acetonitrile in a ratio of 10:90 (v/v) and 1 ml of orthophosphoric acid] with a detection wavelength of 225 nm, a column temperature of 30°C, a flow rate of 1.5 ml/min, and an injection volume of 20 μl. All process, degradant and unknown impurities were separated well with a resolution >2.2 and were estimated accurately without any interference. The recovered values and regression values were 98.7–100.5% and R2 > 0.9999, respectively. The recovery and linearity studies covered the quantitation limit to 150% of the specification limit. The stability‐indicating properties of the developed RP‐HPLC method was assessed from the forced degradation studies. The developed method was successfully applied for real‐time sample analysis of the glipizide dosage form.
Subject
Clinical Biochemistry,Drug Discovery,Pharmacology,Molecular Biology,General Medicine,Biochemistry,Analytical Chemistry
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献