Monte Carlo model of a prototype flat‐panel detector for multi‐energy applications in radiotherapy

Author:

Ozoemelam Ikechi1,Myronakis Marios1,Harris Thomas C.1,Corral Arroyo Pablo2,Huber Pascal2,Jacobson Matthew W.1,Hu Yue‐Houng1,Fueglistaller Rony2,Lehmann Mathias2,Morf Daniel2,Berbeco Ross I.1

Affiliation:

1. Brigham and Women's Hospital Dana‐Farber Cancer Institute and Harvard Medical School Boston Massachusetts USA

2. Varian Imaging Laboratory Baden‐Dattwil Switzerland

Abstract

AbstractBackgroundThe incorporation of multi‐energy capabilities into radiotherapy flat‐panel detectors offers advantages including enhanced soft tissue visualization by reduction of signal from overlapping anatomy such as bone in 2D image projections; creation of virtual monoenergetic images for 3D contrast enhancement, metal artefact reduction and direct acquisition of relative electron density. A novel dual‐layer on‐board imager offering dual energy processing capabilities is being designed. As opposed to other dual‐energy implementation techniques which require separate acquisition with two different x‐ray spectra, the dual‐layer detector design enables simultaneous acquisition of high and low energy images with a single exposure. A computational framework is required to optimize the design parameters and evaluate detector performance for specific clinical applications.PurposeIn this study, we report on the development of a Monte Carlo (MC) model of the imager including model validation.MethodsThe stack‐up of the dual‐layer imager (DLI) was implemented in GEANT4 Application for Tomographic Emission (GATE). The DLI model has an active area of 43×43 cm2, with top and bottom Cesium Iodide (CsI) scintillators of 600 and 800 μm thickness, respectively. Measurement of spatial resolution and imaging of dedicated multi‐material dual‐energy (DE) phantoms were used to validate the model. The modulation transfer function (MTF) of the detector was calculated for a 120 kVp x‐ray spectrum using a 0.5 mm thick tantalum edge rotated by 2.5o. For imaging validation, the DE phantom was imaged using a 140 kVp x‐ray spectrum. For both validation simulations, corresponding measurements were done using an initial prototype of the imager. Agreement between simulations and measurement was assessed using normalized root mean square error (NRMSE) and 1D profile difference for the MTF and phantom images respectively. Further comparison between measurement and simulation was made using virtual monoenergetic images (VMIs) generated from basis material images derived using precomputed look‐up tables.ResultsThe MTF of the bottom layer of the dual‐layer model shows values decreasing more quickly with spatial frequency, compared to the top layer, due to the thicker bottom scintillator thickness and scatter from the top layer. A comparison with measurement shows NRMSE of 0.013 and 0.015 as well as identical MTF50 of 0.8 mm1 and 1.0 mm1 for the top and bottom layer respectively. For the DE imaging of the DE‐phantom, although a maximum deviation of 3.3% is observed for the 10 mm aluminum and Teflon inserts at the top layer, the agreement for all other inserts is less than 2.2% of the measured value at both layers. Material decomposition of simulated scatter‐free DE images gives an average accuracy in PMMA and aluminum composition of 4.9% and 10.3% for 11–30 mm PMMA and 1–10 mm aluminum objects respectively. A comparison of decomposed values using scatter containing measured and simulated DE images shows good agreement within statistical uncertainty.ConclusionValidation using both MTF and phantom imaging shows good agreement between simulation and measurements. With the present configuration of the digital prototype, the model can generate material decomposed images and virtual monoenergetic images.

Funder

National Cancer Institute

Publisher

Wiley

Subject

General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3