Abstract

AbstractBackgroundMany studies have utilized optical camera systems with volumetric scintillators for quality assurances (QA) to estimate the proton beam range. However, previous analytically driven range estimation methods have the difficulty to derive the dose distributions from the scintillation images with quenching and optical effects.PurposeIn this study, a deep learning method utilized to QA was used to predict the beam range and spread‐out Bragg peak (SOBP) for two‐dimensional (2D) map conversion from the scintillation light distribution (LD) into the dose distribution in a water phantom.MethodsThe 2D residual U‐net modeling for deep learning was used to predict the 2D water dose map from a 2D scintillation LD map. Monte Carlo simulations for dataset preparation were performed with varying monoenergetic proton beam energies, field sizes, and beam axis shifts. The LD was reconstructed using photons backpropagated from the aperture as a virtual lens. The SOBP samples were constructed based on monoenergetic dose distributions. The training set, including the validation set, consisted of 8659 image pairs of LD and water dose maps. After training, dose map prediction was performed using a 300 image pair test set generated under random conditions. The pairs of simulated and predicted dose maps were analyzed by Bragg peak fitting and gamma index maps to evaluate the model prediction.ResultThe estimated beam range and SOBP width resolutions were 0.02 and 0.19 mm respectively for varying beam conditions, and the beam range and SOBP width deviations from the reference simulation result were less than 0.1 and 0.8 mm respectively. The simulated and predicted distributions showed good agreement in the gamma analysis, except for rare cases with failed gamma indices in the proximal and field‐marginal regions.ConclusionThe deep learning conversion method using scintillation LDs in an optical camera system with a scintillator is feasible for estimating proton beam range and SOBP width with high accuracy.

Publisher

Wiley

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3