Validity of an ultrasound device to measure bone mineral density

Author:

Puranda Jessica L.1ORCID,Edwards Chris M.1,Weber Vinicius M. R.2,Aboudlal Mohamed1,Semeniuk Kevin1,Adamo Kristi B.1

Affiliation:

1. School of Human Kinetics, Faculty of Health Sciences University of Ottawa Ottawa Canada

2. Laboratory of Experimental and Applied Physiology to Physical Activity Midwest State University of Paraná Guarapuava Paraná Brazil

Abstract

AbstractThis study aims to examine the validity and reliability of the UltraScan650™, a portable ultrasound device, used to measure BMD at the 1/3rd radius position. Fifty‐two female first responders and healthcare providers were assessed using DXA (forearm, femur, lumbar, and total body) and the UltraScan650™. Fat and lean mass were also assessed using the DXA. Pearson correlations, Bland–Altman plots, t‐tests, and linear regressions were used to assess validity. Intra‐class correlation (ICC) coefficients were used to assess reliability. Inter‐rater reliability and repeatability were good (ICC = 0.896 [0.818; 0.942], p < 0.001) and excellent (ICC = 0.917 [0.785; 0.989], p < 0.001), respectively. BMD as measured by the UltraScan650™ was weakly correlated to the DXA (r = 0.382 [0.121; 0.593], p = 0.0052). Bland–Altman plots revealed that the UltraScan650™ underestimated BMD (−0.0569 g/cm2), this was confirmed with a significant paired t‐test (p < 0.001). A linear regression was performed (0.4744 × UltraScan650™ + 0.4170) to provide more information as to the issue of agreement. Bland–Altman plots revealed a negligible bias, supported by a paired t‐test (p = 0.9978). Pearson's correlation revealed a significant relationship (r = −0.771 [−0.862; −0.631], p < 0.0001) between adjusted UltraScan650™—DXA and the average of the two scans (i.e., adjusted UltraScan650™ and DXA), suggesting a proportional constant error and proportional constant variability in measurements of BMD from the UltraScan650™. The UltraScan650™ is not a valid alternative to DXA for diagnostic purposes; however, the UltraScan650™ could be used as a screening tool in the clinical and research setting given the linear transformation is employed.

Funder

Canadian Institutes of Health Research

Coordenação de Aperfeiçoamento de Pessoal de Nível Superior

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3