Affiliation:
1. Department of Radiology Boston University School of Medicine Boston USA
Abstract
AbstractBackgroundTwo‐material decomposition is insufficient to quantify the fat fraction of spinal bone marrow, which is comprised of a mixture of bone minerals, water, and yellow marrow (fat).PurposeTo develop an accurate three‐material decomposition‐based bone marrow fat fraction () quantification technique for dual‐energy CT.MethodsBone marrow edema phantoms containing trabecular bone minerals, water, and fat were constructed using fat fractions and bone mineral density values matching those expected in healthy and edematous bone, and scanned on a commercial dual‐energy CT. Fat quantified by were compared to MRI‐based fat fraction () and conventional two‐material‐decomposition‐based fat fraction () to evaluate its accuracy and dependency on various bone mineral densities.Results demonstrated an excellent correlation with (r = 0.97, R2 = 0.96) in the phantom, significantly more accurate than FF2MD when confounding bone minerals are present (50 mg/cm3: r = 1.02, R2 = 0.95 vs. r = 0.65, R2 = 0.79 (p < 0.01); 100 mg/cm3: r = 0.81, R2 = 0.47 vs. r = 0.21, R2 = 0.21 (p < 0.05)).Conclusions accurately quantified bone marrow fat fraction, when compared with , in the specially constructed bone marrow phantom.