AbstractThe purpose of this study was to investigate the validity of three wrist‐worn devices for estimating energy expenditure (EE) and heart rate (HR) during close “Close Quarter Battle” (CQB). Fifty male soldiers (mean ± SD: age 30.9 ± 4.6 years, height: 1.81 ± 0.64 m and body mass 87.3 ± 7.7 kg) wore three activity monitors (Apple Watch 5, Garmin Forerunner® 935 and GENEActiv accelerometer), a Metamax 3B metabolic cart and a Polar chest strap, whilst conducting a CQB training activity (duration: 26.6 ± 5.0 min). EE and HR data from each test device were compared against criterion measures using ordinary least products regression, 95% limits of agreement, equivalence testing and Mean Absolute Percentage Error (MAPE). Based upon the criterion measure the mean EE for the activity was 372.2 ± 57.6 kcal. All of the devices tested demonstrated fixed and/or proportional bias for EE and a MAPE of >10% (Apple 11.3%, Garmin 15.3%, GENEActiv 57.7%) and therefore did not agree with the criterion. The Apple Watch was a valid method for measuring HR, with a MAPE of 0.6%, and differences with the criterion falling within acceptable limits (≤1 bpm = 83.7%; ≤3 bpm = 97.5% and ≤5 bpm = 97.5%), whereas the Garmin Forerunner® 935 was not valid for measuring HR due to an unacceptable difference compared to the criterion (≤1 bpm = 19.1%; ≤3 bpm = 33.3% and ≤5 bpm = 45.2%). Overall, the Apple Watch 5 can be recommended for measuring HR, but none of the devices are recommended for estimating EE, during CQB.

Funder

Edith Cowan University

Publisher

Wiley

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3