Abstract
Abstract
Two cases are described where, afrer 2 years, a nylon-latex Celestin tube which was left in situ to bypass a benign stricture underwent structural deterioration and produced complications requiring tube replacement. It is recommended that an alternative endo-oesophageal tube should be considered where a lifespan of this order is contemplated.
Publisher
Oxford University Press (OUP)
Reference6 articles.
1. Permanent intubation in inoperable cancer of the oesophagus and cardia;Celestin;Ann. R. Coll. Surg. Engl.,1959
2. Oesophageal intubation in obstructive lesions of the oesophagus;Das;Br. J. Surg.,1973
3. Inoperable carcinoma of the esophagus and cardia. Permanent intubation with the Celestin tube;Fortin;Can. J. Surg.,1965
4. Palliative intubation in the management of esophageal carcinoma;Girardet;Ann. Thoruc. Surg.,1974
5. Celestin tube palliation for malignant esophageal obstruction;Sanfelippo;Surg. Clin. North Am.,1973