Stereotactic radiotherapy of intracranial tumor beds on a ring‐mounted Halcyon LINAC

Author:

Misa Joshua1ORCID,McCarthy Shane1,Clair William St.1,Pokhrel Damodar1ORCID

Affiliation:

1. Medical Physics Graduate Program Department of Radiation Medicine University of Kentucky Lexington Kentucky USA

Abstract

AbstractPurposeThis study sought to evaluate the feasibility and efficacy of the Halcyon Ring Delivery System (RDS) for delivering stereotactic radiotherapy (SRT) treatments for intracranial tumors beds.MethodsTen previously treated brain SRT patients for 30 Gy in five fractions with non‐coplanar HyperArc plans on TrueBeam (6MV‐FFF) were replanned on Halcyon (6MV‐FFF) using the same number of arcs and Eclipse's AcurosXB dose engine. Plan quality evaluation metrics per SRT protocol included: PTV coverage, GTV dose (minimum and mean), target conformity indices (CI), heterogeneity index (HI), gradient index (GI), maximum dose 2 cm away from the PTV (D2cm), and doses to organs‐at‐risk (OAR). Additionally, patient‐specific quality assurance (QA) results and beam‐on‐time (BOT) were analyzed.ResultsThe Halcyon RDS provided highly conformal SRT plans for intracranial tumor beds with similar dose to target. When benchmarked against clinically delivered HyperArc plans, target coverage, CI(s) and HI were statistically similar. The Halcyon plans saw no statistical difference in maximum OAR doses to the brainstem, spinal cord, and cochlea. Due to the machine's coplanar geometry, the Halcyon plans showed a decrease in optic pathway dose (0.75 Gy vs. 2.08 Gy, p = 0.029). Overall, Halcyon's coplanar geometry resulted in a larger GI (3.33 vs. 2.72, p = 0.008) and a larger D2cm (39.59% vs. 29.07%, p < 0.001). In this cohort, multiple cases had the PTV and the optic pathway in the same axial plane. In one such instance, the PTV was <2 cm away from the optic pathway but even at this close proximity OAR, Halcyon still adequately spared the optic pathway. Additionally, the Halcyon's geometry provided slightly larger amount of normal brain dose receiving 24.4 Gy (8.99 cc vs. 7.36 cc) and 28.8 Gy (2.9 cc vs. 2.5 cc), although statistically insignificant. The Halcyon plans achieved similar delivery accuracy, quantified by patient‐specific QA results evaluated with a 2%/2 mm gamma criteria (99.42% vs. 99.70%). For both plans, independent Monte Carlo second checks calculation agreed within 1%. Average Halcyon BOT was slightly higher by 0.35 min (p = 0.045), however, due to the one‐step patient set‐up and verification overall estimated treatment times on Halcyon were lower compared to HyperArc treatments (7.61 min vs. 10.26 min, p < 0.001).ConclusionsWhen benchmarked against clinically delivered HyperArc treatments, the Halcyon brain SRT plans provided similar plan quality and delivery accuracy but achieved faster overall treatment times. We have started treating select brain SRT patients on the Halcyon RDS for patients having tumor beds greater than 1 cm in diameter with the closest OAR distance of greater than 2 cm away from the target. We recommend other clinics to consider commissioning SRT treatments on their Halcyon systems–allowing including remote Halcyon‐only clinics to provide exceptionally high‐quality therapeutic brain SRT treatments to an otherwise underserved patient cohort.

Publisher

Wiley

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3