Dose‐rate dependence and IMRT QA suitability of EBT3 radiochromic films for pulse reduced dose‐rate radiotherapy (PRDR) dosimetry

Author:

Khan Ahtesham Ullah1,Radtke Jeff1,Hammer Clifford1,Malyshev Julia1,Morris Brett2,Glide‐Hurst Carri12,DeWerd Larry1,Culberson Wesley1,Bayliss Adam2

Affiliation:

1. Department of Medical Physics, School of Medicine and Public Health University of Wisconsin‐Madison Madison Wisconsin USA

2. Department of Human Oncology, School of Medicine and Public Health University of Wisconsin‐Madison Madison Wisconsin USA

Abstract

AbstractBackgroundPulsed reduced dose rate (PRDR) is an emerging radiotherapy technique for recurrent diseases. It is pertinent that the linac beam characteristics are evaluated for PRDR dose rates and a suitable dosimeter is employed for IMRT QA.PurposeThis study sought to investigate the pulse characteristics of a 6 MV photon beam during PRDR irradiations on a commercial linac. The feasibility of using EBT3 radiochromic film for use in IMRT QA was also investigated by comparing its response to a commercial diode array phantom.MethodsA plastic scintillator detector was employed to measure the photon pulse characteristics across nominal repetition rates (NRRs) in the 5–600 MU/min range. Film was irradiated with dose rates in the 0.033–4 Gy/min range to study the dose rate dependence. Five clinical PRDR treatment plans were selected for IMRT QA with the Delta4 phantom and EBT3 film sheets. The planned and measured dose were compared using gamma analysis with a criterion of 3%/3 mm. EBT3 film QA was performed using a cumulative technique and a weighting factor technique.ResultsNegligible differences were observed in the pulse width and height data between the investigated NRRs. The pulse width was measured to be 3.15 ± 0.01 and the PRF was calculated to be 3–357 Hz for the 5–600 MU/min NRRs. The EBT3 film was found to be dose rate independent within 3%. The gamma pass rates (GPRs) were above 99% and 90% for the Delta4 phantom and the EBT3 film using the cumulative QA method, respectively. GPRs as low as 80% were noted for the weighting factor EBT3 QA method.ConclusionsAltering the NRRs changes the mean dose rate while the instantaneous dose rate remains constant. The EBT3 film was found to be suitable for PRDR dosimetry and IMRT QA with minimal dose rate dependence.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3