Impurity Analysis of 5.5 N Nitrogen Gas and Its Impact on the Preparation of Class I Type Calibration Gas Mixtures

Author:

Garg Gazal12,Kumari Poonam12,Soni Daya12ORCID,Singh Khem1,Aggarwal Shankar G12,Deep Akash23

Affiliation:

1. CSIR-National Physical Laboratory Dr. K. S. Krishnan Marg New Delhi 110012 India

2. Academy of Scientific and Innovative Research (AcSIR) Ghaziabad 201002 India

3. CSIR-Central Scientific Instruments Organization, Sector 30 C Chandigarh 160030 India

Abstract

AbstractKnowledge of impurities in matrix gas is one of the critical requirements for the preparation of class I type calibration gas mixtures (primary reference gas mixtures, PRGMs) as per ISO 6142‐1. PRGMs are prepared gravimetrically using high‐purity component gas and high‐grade nitrogen, here 99.9995 % (5.5 N) as matrix gas. In this study, quantification of gaseous impurities such as moisture (H2O), methane (CH4), and carbon monoxide (CO) in 5.5 N nitrogen is performed using CRDS (cavity ring down spectroscopy). Carbon dioxide (CO2) and total hydrocarbons (THC) are analyzed using gas chromatography flame ionization detector (GC‐FID), while the determination of nitrogen dioxide (NO2) impurity is conducted colorimetrically using a UV‐Vis spectrophotometer. Total impurities in 5.5 N nitrogen are found to be (2945±21.5) ppb. The impact of utilizing this nitrogen is examined through the stability study of ambient range binary component gas mixtures of CH4 and CO in nitrogen and found that 5.5 N nitrogen can be used as matrix gas for preparing PRGMs of CH4 and CO. However, the component gas impurity present in nitrogen potentially affect the accuracy of gravimetrically prepared gas mixture.

Publisher

Wiley

Subject

General Chemistry

Reference32 articles.

1. S. Petrozzi Practical Instrumental Analysis: Methods Quality Assurance and Laboratory Management 2012 Ist Edition.

2. M. W. Sayers T. D. Gillespie C. A. V. Queiroz Establishing Correlation and a Calibration Standard for Measurements 1986 WTP-45.

3. ISO 19229 : 2019 Gas analysis - Purity Analysis and the treatment of purity data Available from:https://www.iso.org/standard/72010.html.

4. ISO 6142–1 : 2015 Gas analysis - Preparation of Calibration Gas Mixtures Available from:https://doi.org/10.3403/30419855.

5. Revision of ISO 19229 to support the certification of calibration gases for purity