Technical note: Characterization, validation, and spectral optimization of a dedicated breast CT system for contrast‐enhanced imaging

Author:

Pautasso Juan J.1,Michielsen Koen1,Sechopoulos Ioannis123

Affiliation:

1. Department of Medical Imaging Radboud University Medical Center Nijmegen The Netherlands

2. Dutch Expert Centre for Screening (LRCB) Nijmegen The Netherlands

3. Technical Medical Centre University of Twente Enschede The Netherlands

Abstract

AbstractBackgroundThe development of a new imaging modality, such as 4D dynamic contrast‐enhanced dedicated breast CT (4D DCE‐bCT), requires optimization of the acquisition technique, particularly within the 2D contrast‐enhanced imaging modality. Given the extensive parameter space, cascade‐systems analysis is commonly used for such optimization.PurposeTo implement and validate a parallel‐cascaded model for bCT, focusing on optimizing and characterizing system performance in the projection domain to enhance the quality of input data for image reconstruction.MethodsA parallel‐cascaded system model of a state‐of‐the‐art bCT system was developed and model predictions of the presampled modulation transfer function (MTF) and the normalized noise power spectrum (NNPS) were compared with empirical data collected in the projection domain. Validation was performed using the default settings of 49 kV with 1.5 mm aluminum filter and at 65 kV and 0.257 mm copper filter. A 10 mm aluminum plate was added to replicate the breast attenuation. Air kerma at the isocenter was measured at different tube current levels. Discrepancies between the measured projection domain metrics and model‐predicted values were quantified using percentage error and coefficient of variation (CoV) for MTF and NNPS, respectively. The optimal filtration was for a 5 mm iodine disk detection task at 49, 55, 60, and 65 kV. The detectability index was calculated for the default aluminum filtration and for copper thicknesses ranging from 0.05 to 0.4 mm.ResultsAt 49 kV, MTF errors were +5.1% and −5.1% at 1 and 2 cycles/mm, respectively; NNPS CoV was 5.3% (min = 3.7%; max = 8.5%). At 65 kV, MTF errors were ‐0.8% and ‐3.2%; NNPS CoV was 13.1% (min = 11.4%; max = 16.9%). Air kerma output was linear, with 11.67 µGy/mA (R= 0.993) and 19.14 µGy/mA (R= 0.996) at 49 and 65 kV, respectively. For iodine detection, a 0.25 mm‐thick copper filter at 65 kV was found optimal, outperforming the default technique by 90%.ConclusionThe model accurately predicts bCT system performance, specifically in the projection domain, under varied imaging conditions, potentially contributing to the enhancement of 2D contrast‐enhanced imaging in 4D DCE‐bCT.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3