Affiliation:
1. McConnell Brain Imaging Centre Montreal Neurological Institute and Hospital Montreal QC Canada
2. Department of Biomedical Engineering McGill University Montreal QC Canada
3. Department of Neurology and Neurosurgery McGill University Montreal QC Canada
4. Siemens Healthcare Limited Montreal QC Canada
Abstract
AbstractPhosphorus (31P) magnetic resonance spectroscopic imaging (MRSI) can serve as a critical tool for more direct quantification of brain energy metabolism, tissue pH, and cell membrane turnover. However, the low concentration of 31P metabolites in biological tissue may result in low signal‐to‐noise ratio (SNR) in 31P MRS images. In this work, we present an innovative design and construction of a 31P radiofrequency coil for whole‐brain MRSI at 7 T. Our coil builds on current literature in ultra‐high field 31P coil design and offers complete coverage of the brain, including the cerebellum and brainstem. The coil consists of an actively detunable volume transmit (Tx) resonator and a custom 24‐channel receive (Rx) array. The volume Tx resonator is a 16‐rung high‐pass birdcage coil. The Rx coil consists of a 24‐element phased array composed of catered loop shapes and sizes built onto a custom, close‐fitting, head‐shaped housing. The Rx array was designed to provide complete coverage of the head, while minimizing mutual coupling. The Rx configuration had a mean
reflection coefficient better than −20 decibels (dB) when the coil was loaded with a human head. The mean mutual coupling (
) among Rx elements, when loaded with a human head, was −16 dB. In phantom imaging, the phased array produced a central SNR that was 4.4‐fold higher than the corresponding central SNR when operating the 31P birdcage as a transceiver. The peripheral SNR was 12‐fold higher when applying the optimized phased array. In vivo 3D 31P MRSI experiments produced high‐quality spectra in the cerebrum gray and white matter, as well as in the cerebellum. Characteristic phosphorus metabolites related to adenosine triphosphate metabolism and cell membrane turnover were distinguishable across all brain regions. In summary, our results demonstrate the potential of our novel coil for accurate, whole‐brain 31P metabolite quantification.
Funder
Canada Foundation for Innovation
Natural Sciences and Engineering Research Council of Canada
Canada First Research Excellence Fund