A rotating beam‐blocker method for cone beam CT scatter correction

Author:

Cui Hehe1234,Zhan Haolin1,Yuan Xiaogang1,Yang Yidong156

Affiliation:

1. Department of Engineering and Applied Physics University of Science and Technology of China Hefei Anhui People's Republic of China

2. Cancer Center Union Hospital Tongji Medical College Huazhong University of Science and Technology Wuhan People's Republic of China

3. Institute of Radiation Oncology Union Hospital Tongji Medical College Huazhong University of Science and Technology Wuhan People's Republic of China

4. Hubei Key Laboratory of Precision Radiation Oncology Union Hospital Tongji Medical College Huazhong University of Science and Technology Wuhan People's Republic of China

5. Department of Radiation Oncology the First Affiliated Hospital of USTC Division of Life Sciences and Medicine University of Science and Technology of China Hefei Anhui People's Republic of China

6. Ion Medical Research Institute University of Science and Technology of China Hefei Anhui People's Republic of China

Abstract

AbstractBackgroundCone beam CT (CBCT) is widely utilized in clinics. However, the scatter artifact degrades the CBCT image quality, hampering the expansion of CBCT applications. Recently, beam‐blocker methods have been used for CBCT scatter correction and proved their high cost‐effectiveness.PurposeA rotating beam‐blocker (RBB) method for CBCT scatter correction was proposed to complete scatter correction and image reconstruction within a single scan in both full‐ and half‐fan scan scenarios.MethodsThe RBB consisted of two open regions and two blocked regions, and was designed as a centrosymmetric structure. The open and blocked projections could be alternatively obtained within one single rotation. The open projections were corrected with the scatter signal calculated from the blocked projections, and then used to reconstruct the 3D image via the Feldkamp‐Davis‐Kress algorithm. The performance of the RBB method was evaluated on head and pelvis phantoms in scenarios with and without a bowtie filter. The images obtained from nine repeated scans in each scenario were used to calculate the evaluation metrics including the CT number error, spatial nonuniformity (SNU) and contrast‐to‐noise ratio (CNR).ResultsFor the head phantom, the CT number error was decreased to <5 after scatter correction from >200 HU before correction when scanned without a bowtie filter, and to <4 from >160 HU when scanned with a full bowtie filter. For the pelvis phantom, the CT number error was reduced to <12 after scatter correction from >250 HU before correction when scanned without a bowtie filter, and to <10 from >190 HU when scanned with a half bowtie filter. After scatter correction, the uniformity and contrast were both improved, resulting in an SNU of >79% decrease and CNR of >2 times increase, respectively.ConclusionsHigh‐quality CBCT images could be obtained in a single scan after using the proposed RBB method for scatter correction, enabling more accurate image guidance for surgery and radiation therapy applications. With almost no time delay between the successive open and blocked projections, the RBB method could eliminate the motion‐induced anatomical mismatches between the corresponding open and blocked projections and could find particular usefulness in thoracic and abdominal imaging.

Funder

Fundamental Research Funds for the Central Universities

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3