Development and clinical application of a GPU‐based Monte Carlo dose verification module and software for 1.5 T MR‐LINAC

Author:

Cheng Bo1,Xu Yuan2,Li Shijun1,Ren Qiang3,Pei Xi13,Men Kuo2,Dai Jianrong2,Xu Xie George14

Affiliation:

1. School of Nuclear Science and Technology University of Science and Technology of China Hefei China

2. Department of Radiation Oncology National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital Chinese Academy of Medical Sciences and Peking Union Medical College Beijing China

3. Technology Development Department Anhui Wisdom Technology Company Limited Hefei China

4. Department of Radiation Oncology The First Affiliated Hospital of University of Science and Technology of China Hefei China

Abstract

AbstractBackgroundAdaptive radiotherapy (ART) has made significant advances owing to magnetic resonance linear accelerator (MR‐LINAC), which provides superior soft‐tissue contrast, fast speed and rich functional magnetic resonance imaging (MRI) to guide radiotherapy. Independent dose verification plays a critical role in discovering errors, while several challenges remain in MR‐LINAC.PurposeA Monte Carlo‐based GPU‐accelerated dose verification module for Unity is proposed and integrated into the commercial software ArcherQA to achieve fast and accurate quality assurance (QA) for online ART.MethodsElectron or positron motion in a magnetic field was implemented, and a material‐dependent step‐length limit method was used to trade off speed and accuracy. Transport was verified by dose comparison with EGSnrc in three A‐B‐A phantoms. Then, an accurate Monte Carlo‐based Unity machine model was built in ArcherQA, including an MR‐LINAC head, cryostat, coils, and treatment couch. In particular, a mixed model combining measured attenuation and homogeneous geometry was adopted for the cryostat. Several parameters in the LINAC model were tuned to commission it in the water tank. An alternating open‐closed MLC plan on solid water measured with EBT‐XD film was used to verify the LINAC model. Finally, the ArcherQA dose was compared with ArcCHECK measurements and GPUMCD in 30 clinical cases through the gamma test.ResultsArcherQA and EGSnrc were well matched in three A‐B‐A phantom tests, and the relative dose difference (RDD) was less than 1.6% in the homogenous region. A Unity model was commissioned in the water tank, and the RDD in the homogenous region was less than 2%. In the alternating open‐closed MLC plan, the gamma result (3%/3 mm) between ArcherQA and Film was 96.55%, better than the gamma result between GPUMCD and Film (92.13%). In 30 clinical cases, the mean three‐dimensional (3D) gamma result (3%/2 mm) was 99.36% ± 1.28% between ArcherQA and ArcCHECK for the QA plans and 99.27% ± 1.04% between ArcherQA and GPUMCD for the clinical patient plans. The average dose calculation time was 106 s in all clinical patient plans.ConclusionsA GPU‐accelerated Monte Carlo‐based dose verification module was developed and built for the Unity MR‐LINAC. The fast speed and high accuracy were proven by comparison with EGSnrc, commission data, the ArcCHECK measurement dose, and the GPUMCD dose. This module can achieve fast and accurate independent dose verification for Unity.

Funder

National Natural Science Foundation of China

Publisher

Wiley

Subject

General Medicine